The
View
menu offers you two choices:
1.
Options
•
Selecting this option brings up the COM
setting form.
•
See the “COM” icon, and description below,
for an explanation of how to set the COM
options.
2.
Keypad
•
Selecting this option toggles the graphic
pendant on and off.
•
The graphic pendant is displayed when a
checkmark (
a
) appears beside the selection.
Choose CPU Menu
This function allows the user to operate the service
software either through CPU1 or CPU2. This does
not change the CPU that the primary hand pendant
uses.
Load Data Menu
This option allows the user to read the default setting,
the settings programmed in CPU1, or the settings
programmed in CPU2.
Selecting this option loads the default data (i.e.,
parameter settings) into the active CPU.
Free Access Menu
This option prompts the user for a “Key-Code” for
virtual control of table functions. This is a purchased
feature, please see your BERCHTOLD Corporation
Sales/Service Representative.
Open File Button
Clicking the Open File button results presents a
standard Windows Open File screen (Figure 6-7).
FIGURE 6-7.
WINDOWS FILE OPEN
SELECTION SCREEN
Save File Button
Clicking the Save File button results presents a
standard Windows File Save screen (Figure 6-8).
FIGURE 6-8.
WINDOWS FILE SAVE
SCREEN
Copy to Button
The copy to button is inactive in the customer and
field service versions of the service program.
Factory technicians use the Copy to button to copy
the contents of the current CPU to the other CPU.
You should use this function any time you make any
change in the table’s programming or settings. It is
important that both CPUs contain the same operating
instructions so that the operators always know what
to expect while using the table.
Содержание OPERON B 810
Страница 1: ...OPERON B 810 SERVICE PARTS MANUAL ...
Страница 65: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 2 Rev 1 70 000 0027 FIGURE 5 1 C8 HYDRAULIC SCHEMATIC ...
Страница 68: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 70 000 0027 Rev 1 5 5 WIRING DIAGRAM SHEET 2 DETAIL ...
Страница 69: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 6 Rev 1 70 000 0027 WIRING DIAGRAM SHEET 3 DETAIL ...
Страница 70: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 70 000 0027 Rev 1 5 7 WIRING DIAGRAM SHEET 4 DETAIL ...
Страница 71: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 8 Rev 1 70 000 0027 BASE WIRING DIAGRAM ...
Страница 104: ...OPERON B 810 SURGICAL TABLE PARTS 7 4 Rev 1 70 000 0027 THIS PAGE INTENTIONALLY BLANK ...
Страница 109: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 5 ...
Страница 110: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 6 ...
Страница 116: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 12 BERCHTOLD Corp BERCHTOLD Corp ...
Страница 117: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 13 B E R C H T O L D C o r p ...
Страница 118: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 14 BERCHTOLD Corp BERCHTOLD Corp ...
Страница 120: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 16 BERCHTOLD Corp BERCHTOLD Corp ...