OPERON B 810 SURGICAL TABLE
SPECIAL SAFETY INSTRUCTIONS
70-000-0027
Rev.
1
S - 5
life. (See drawing 306762 for correct
fluid designation.)
ELECTROMAGNETIC INTERFERENCE
WARNING
WARNING
— When using the table
in the vicinity of high-frequency
surgical equipment, such as cardiac
defibrillators and cardiac
defibrillator monitors, refer to the
equipment manufacturers’
instructions to ensure compatibility
issues are considered.
ELECTRICAL SYSTEM
WARNING
WARNING –
When servicing the
table, turn power off or remove the
appropriate fuses to prevent personal
injury. If power is required, use care
to avoid contact with live parts.
WARNING
WARNING –
Even a “dead” battery
can contain enough power to cause a
serious shock or explosive spark.
Always disconnect the battery charger
plug at the power center before
disconnecting cables.
WARNING
WARNING –
The batteries contain
acid, which can cause serious burns.
While the batteries are sealed, a dead
battery may be damaged and may leak.
Always handle batteries in a way that
will prevent acid from spilling on your
skin or clothes.
SERVICE PROGRAM
CAUTION
CAUTION —
The Service Program
will allow you to set motor parameters
outside the min. and max. values of
100 to 1000. Setting the parameters to
the extreme min. or max. value could
lead to damage to the table or its
components, including a failure to
operate. Always double check your
settings before clicking on “Save
Parameters”.
CAUTION
CAUTION —
The Service Program
will allow you to set valve parameters
outside the min and max range of 100-
300MSec. This could lead to damage
to the table or its components,
including a failure to operate. Always
double check your settings before
clicking on “Save Parameters”.
CAUTION
CAUTION —
The Service Program
will allow you to set sensor parameters
outside the min and max range shown
in TABLE 6-1. This could lead to
damage to the table or its components,
including a failure to operate. Always
double check your settings before
clicking on “Save Parameters”.
CAUTION
CAUTION —
The Service Program
will allow you to set ACCU (battery)
parameters outside the ranges shown
in TABLE 6-2. This could lead to false
readings on the (physical and graphic)
pendant LEDs, including a failure to
operate. Always double check your
settings before clicking on “Save
Parameters”.
Содержание OPERON B 810
Страница 1: ...OPERON B 810 SERVICE PARTS MANUAL ...
Страница 65: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 2 Rev 1 70 000 0027 FIGURE 5 1 C8 HYDRAULIC SCHEMATIC ...
Страница 68: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 70 000 0027 Rev 1 5 5 WIRING DIAGRAM SHEET 2 DETAIL ...
Страница 69: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 6 Rev 1 70 000 0027 WIRING DIAGRAM SHEET 3 DETAIL ...
Страница 70: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 70 000 0027 Rev 1 5 7 WIRING DIAGRAM SHEET 4 DETAIL ...
Страница 71: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 8 Rev 1 70 000 0027 BASE WIRING DIAGRAM ...
Страница 104: ...OPERON B 810 SURGICAL TABLE PARTS 7 4 Rev 1 70 000 0027 THIS PAGE INTENTIONALLY BLANK ...
Страница 109: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 5 ...
Страница 110: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 6 ...
Страница 116: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 12 BERCHTOLD Corp BERCHTOLD Corp ...
Страница 117: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 13 B E R C H T O L D C o r p ...
Страница 118: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 14 BERCHTOLD Corp BERCHTOLD Corp ...
Страница 120: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 16 BERCHTOLD Corp BERCHTOLD Corp ...