OPERON B 810 SURGICAL TABLE
SERVICE PROGRAM
6-10 Rev.
1
70-000-0027
Valve Tab
The valve tab allows you to control the amount of
time that the hydraulic valves open before the motor
starts pumping fluid (Figure 6-13). Opening the
valves too early could allow the table to move a
perceptible amount before the motor activates to
provide positive control over the movement.
The following delay can be set for each movement:
•
Start Delay (msec)
This is the time the processor waits after opening
the valve before starting the hydraulic motor.
•
Stop Delay (msec)
This is the time the processor waits to close the
valve after the hydraulic motor has stopped.
Motor Power settings can be adjusted in the range of
15-100%.
CAUTION
CAUTION —
The Service Program will
allow you to set valve parameters
outside the min and max range of 100-
300MSec. This could lead to damage to
the table or its components, including a
failure to operate. Always double check
your settings before clicking on “Save
Parameters”.
To set the value of the parameter, use your cursor to
highlight the number you wish to change. Type in
the new value to replace the old value.
After setting all of the parameters to the desired
levels, click on the “Save parameters” button to
record them in the table’s CPU.
FIGURE 6-13.
VALVE SETTINGS FORM
Sensor Tab
The sensor tab allows you to set the parameters for
the following sensors (Figure 6-14):
•
Tilt
•
Trendelenburg
•
Back
•
Leg
•
Slide
Sensors for kidney movements are mechanical
devices that are adjusted physically.
CAUTION
CAUTION —
The Service Program will
allow you to set sensor parameters
outside the min and max range shown in
TABLE 6-1. This could lead to damage
to the table or its components, including
Содержание OPERON B 810
Страница 1: ...OPERON B 810 SERVICE PARTS MANUAL ...
Страница 65: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 2 Rev 1 70 000 0027 FIGURE 5 1 C8 HYDRAULIC SCHEMATIC ...
Страница 68: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 70 000 0027 Rev 1 5 5 WIRING DIAGRAM SHEET 2 DETAIL ...
Страница 69: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 6 Rev 1 70 000 0027 WIRING DIAGRAM SHEET 3 DETAIL ...
Страница 70: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 70 000 0027 Rev 1 5 7 WIRING DIAGRAM SHEET 4 DETAIL ...
Страница 71: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 8 Rev 1 70 000 0027 BASE WIRING DIAGRAM ...
Страница 104: ...OPERON B 810 SURGICAL TABLE PARTS 7 4 Rev 1 70 000 0027 THIS PAGE INTENTIONALLY BLANK ...
Страница 109: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 5 ...
Страница 110: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 6 ...
Страница 116: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 12 BERCHTOLD Corp BERCHTOLD Corp ...
Страница 117: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 13 B E R C H T O L D C o r p ...
Страница 118: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 14 BERCHTOLD Corp BERCHTOLD Corp ...
Страница 120: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 16 BERCHTOLD Corp BERCHTOLD Corp ...