OPERON B 810 SURGICAL TABLE
GENERAL INFORMATION
70-000-0027
Rev.
1
1-1
GENERAL INFORMATION
GENERAL DESCRIPTION
The Berchtold
OPERON B 810
surgical table is a
remote controlled, full-function design offering a
high level of patient positioning and imaging
flexibility to the surgical team. The table can be
activated by AC power through the power cord (input
range of 115V ±10%) or by battery power, which
requires no power cord except when charging the
batteries.
An electric motor with a belt driven screw is used to
slide the table top head to foot.
Hydraulic cylinders are used to articulate the table's
sections. Hydraulic pressure is provided by a motor
driven pump during normal use, and by a manual foot
pump in manual back-up mode.
A hand-held control pendant is used to control table
positions.
The tabletop is divided into the following sections
(FIGURE 1-1):
•
Headrest
•
Back Section
•
Kidney Elevator
•
Seat Section
•
Leg Section
A leg extension can be added to extend the tabletop
surface.
Leg Section
Seat
Section
Primary
Hand
Pendant
Floor Feet
Primary Hand
Pendant
Receptacle
Footswitch
Receptacle
Manual Foot
Pump Pedal
Located on this
end of Base:
Power Entry
Module, Fuses,
and Auxiliary
Pendant Drawer
Back
Section
Headrest
Kidney
Elevator
FIGURE 1-1
OPERON B 810 SURGICAL TABLE
RADIOLUCENT TABLE TOP AND PADS
CAUTION
CAUTION
— The anti-static
properties of the table depend on the
use of the recommended mattress.
Compromise of the anti-static
properties could expose the patient or
OR personnel to an electrostatic
shock. It could also deliver a shock to
the table’s electronics that could
damage or destroy critical
components.
The tabletop surface is made from electrically
conductive radiolucent panels. The table is fitted with
a matching set of conductive radiolucent pads. Velcro
tape holds the pads in place.
CAUTION
CAUTION
— Protect the top and
bottom surfaces from nicks and
scratches. Scratches more than 0.005
inches (127 microns) deep may appear
on x-rays. Damaged or scratched
panels should be replaced if x-ray
quality is affected.
Each table section has a cassette channel that can
accommodate x-ray cassettes up to 14'' x 17'' (356 x
432 mm).
INTERCHANGEABLE HEADREST
The tabletop has perineal cutouts in both seat and
back sections. The back/seat pad can be reversed to
expose the desired cutout. The headrest can be
interchanged to allow for normal and reversed patient
orientation.
Содержание OPERON B 810
Страница 1: ...OPERON B 810 SERVICE PARTS MANUAL ...
Страница 65: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 2 Rev 1 70 000 0027 FIGURE 5 1 C8 HYDRAULIC SCHEMATIC ...
Страница 68: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 70 000 0027 Rev 1 5 5 WIRING DIAGRAM SHEET 2 DETAIL ...
Страница 69: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 6 Rev 1 70 000 0027 WIRING DIAGRAM SHEET 3 DETAIL ...
Страница 70: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 70 000 0027 Rev 1 5 7 WIRING DIAGRAM SHEET 4 DETAIL ...
Страница 71: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 8 Rev 1 70 000 0027 BASE WIRING DIAGRAM ...
Страница 104: ...OPERON B 810 SURGICAL TABLE PARTS 7 4 Rev 1 70 000 0027 THIS PAGE INTENTIONALLY BLANK ...
Страница 109: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 5 ...
Страница 110: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 6 ...
Страница 116: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 12 BERCHTOLD Corp BERCHTOLD Corp ...
Страница 117: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 13 B E R C H T O L D C o r p ...
Страница 118: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 14 BERCHTOLD Corp BERCHTOLD Corp ...
Страница 120: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 16 BERCHTOLD Corp BERCHTOLD Corp ...