OPERON B 810 SURGICAL TABLE
SERVICE PROGRAM
6-24 Rev.
1
70-000-0027
No.
1
Error
Reaction Description
Solution
50
Valve-14 fault
(KIDNEY UP)
Continue Valve disconnected, wiring broken, short
circuit, damaged solenoid coil
•
Connect valve.
•
Identify and replace
broken wiring.
•
Locate and remove
cause of short circuit.
•
Replace the Solenoid
coil.
53
Hand Pendant fault Continue Push button defective during initial test,
wiring broken
NOTE: This error code will be recorded if
the operator presses two or more buttons
when switching the pendant on from “sleep”
mode.
•
Check connection
•
Check all pendant
functions. Replace
pendant if functions
to not operate as
expected.
55
Foot Control Panel
fault
Continue Pedal pressed or defect during initial test
NOTE: This error code will be recorded if
the operator presses two or more buttons
when switching the foot control on from
“sleep” mode.
•
Check connection
•
Check all footswitch
functions. Replace
footswitch if
functions do not
operate as expected.
58
Event:
Communication
command “Reset”
Reset
Micro controller was reset by communication
command "Reset"
No action required.
59
Reset Instruction
Executed
Reset
Micro controller was reset by software
instruction
No action required.
60
Watchdog timeout
reset
Reset
Oscillator / Crystal 10 MHz defective,
incorrect frequency
•
Non-critical error.
•
Run table through all
functions to ensure
proper operation. If
functions fail to
operate as expected,
replace the CPU.
61 Brown-out
reset Reset
Supply voltage for the Micro controller
(n5V) is between 4.2 volt and 4.6
volt. Battery almost discharged, Voltage
Regulator defective
•
Charge battery
•
Replace charging
circuit
62
Stack
overflow/underflow
Reset
Runtime error in the software caused by
inadequate program memory
Replace the CPU
63
Unknown RESET
condition
Reset
Micro controller was reset through unknown
cause
NOTE: This error can be caused by
electromagnetic radiation strong enough to
penetrate the table’s shielding.
No action required.
If problem persists
check for EMI source
64
Event: Power Up
(First time ON)
Continue This is not an error. It is a counter that
records the "Power Up" events.
No action required.
1 Missing numbers in the sequence indicate error codes reserved for future releases. If you receive an unnumbered error code, contact
Berchtold Technical Service for assistance.
Содержание OPERON B 810
Страница 1: ...OPERON B 810 SERVICE PARTS MANUAL ...
Страница 65: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 2 Rev 1 70 000 0027 FIGURE 5 1 C8 HYDRAULIC SCHEMATIC ...
Страница 68: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 70 000 0027 Rev 1 5 5 WIRING DIAGRAM SHEET 2 DETAIL ...
Страница 69: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 6 Rev 1 70 000 0027 WIRING DIAGRAM SHEET 3 DETAIL ...
Страница 70: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 70 000 0027 Rev 1 5 7 WIRING DIAGRAM SHEET 4 DETAIL ...
Страница 71: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 8 Rev 1 70 000 0027 BASE WIRING DIAGRAM ...
Страница 104: ...OPERON B 810 SURGICAL TABLE PARTS 7 4 Rev 1 70 000 0027 THIS PAGE INTENTIONALLY BLANK ...
Страница 109: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 5 ...
Страница 110: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 6 ...
Страница 116: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 12 BERCHTOLD Corp BERCHTOLD Corp ...
Страница 117: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 13 B E R C H T O L D C o r p ...
Страница 118: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 14 BERCHTOLD Corp BERCHTOLD Corp ...
Страница 120: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 16 BERCHTOLD Corp BERCHTOLD Corp ...