Options and accessories
88
interface must be isolated using an appropriate device.
●
Only connect printers which use an IBM 8-bit character set. Set the printer up so that it runs in
IBM emulation.
8.4 ST6180 barcode reading wand, DLC7070 barcode scanner
The ST6180 barcode reading wand or DLC7070 scanner can be used to read master data,
such as the device ID, type name or test engineer's name.
Devices and types can be selected in the relevant catalogue with the barcode scanner. When
a barcode is read in the main menu, UNIMET® 1100ST will attempt to find a device with this
serial number in its device catalogue. If it finds such a device, the test can be started directly.
If there are a number of different devices in the device catalogue with the same serial
number, the device found first is displayed.
How to activate the barcode scanner
●
Connect the barcode reading wand or barcode scanner to the RS-232 serial interface on the
Unimet.
●
Select the appropriate barcode scanner in the system settings (see “Barcode reading” on
page 25). The barcode reader is supplied set up for use with UNIMET® 1100ST.
●
Please observe the information in the manual supplied with the barcode scanner.
●
Test the function of the barcode scanner. Barcodes on merchandise can be used for this
purpose.
●
If you create your own barcodes:
– Use one of the barcode types indicated in the barcode scanner manual.
– Use test printouts to check whether the print quality is sufficient to ensure reliable reading.
Important information about the ST6180 barcode reading wand
●
Following logon and every time the Unimet is started up, the ST6180 barcode reading wand
will need about 20 seconds before it is ready for operation. It may need longer the first time it is
used.
●
The red LED in the tip of the reading wand will flash to indicate readiness for operation.
●
A short beep will indicate that the barcode has been read successfully.
●
When the ST6180 barcode reading wand is not in use, no objects should be positioned in front
of the tip of the wand. Otherwise the reading wand will attempt to read without success; a beep
will sound as long as the object remains in front of the tip of the reading wand.
8.5 VK701 adapter
The VK701 adapter enables UNIMET® 1100ST to check extension cables and device power
supply cables. This automated test includes checking that the wiring is correct and verifying
the connection between the individual pins on the plug and socket. The insulation resistance
of the wires in relation to the PE conductor is also tested.
Содержание UNIMET 1000ST
Страница 50: ...Operation and setting 50 TGH1256en 10 2005...
Страница 80: ...PC compatible functions of the UNIMET 1100ST 80 TGH1256en 10 2005...
Страница 82: ...Maintenance and calibration 82 TGH1256en 10 2005...
Страница 90: ...Options and accessories 90 TGH1256en 10 2005...
Страница 96: ...Data 96 TGH1256en 10 2005...
Страница 99: ......