Delete all data
[-> Main menu -> F5: System administration -> F3: Database maintenance… -> F3: Catalogues
-> F3: Delete all data]
This function is used for example if the UNIMET® 1100ST is used by different departments
and each department transfers its data for periodic testing from PC to the UNIMET® 1100ST.
Repair database
[-> Main menu -> F5: System administration -> F3: Database maintenance… -> F3: Catalogues
-> F4: Repair database]
If individual data records have become unusable (e.g. due to a power failure during the test
sequence), the situation can be rectified using function "F4: Repair database".
4.5.5
System info
[-> Main menu -> F5: System administration -> F4: System info]
The menu branches to the system self test screen. The screen contains information about the
software, firmware and hardware versions as well as the type of test probe selected. Keep
this information to hand should you need to contact BENTRON
®
Service for assistance.
As a new empty database is created, this action frees up memory. Existing
database problems may be deleted along with the data. An additional prompt
will prevent unintentional deletion of the database.
The process takes several minutes and must not be interrupted.
The UNIMET® 1000ST PCMCIA card must not be removed whilst repairs are in
progress. Removing the card may render the Unimet database unusable.
Tip: You can also use the UNIData1100 PC software to repair the database.
Содержание UNIMET 1000ST
Страница 50: ...Operation and setting 50 TGH1256en 10 2005...
Страница 80: ...PC compatible functions of the UNIMET 1100ST 80 TGH1256en 10 2005...
Страница 82: ...Maintenance and calibration 82 TGH1256en 10 2005...
Страница 90: ...Options and accessories 90 TGH1256en 10 2005...
Страница 96: ...Data 96 TGH1256en 10 2005...
Страница 99: ......