Operation and setting
39
4.5.5.2
System self test with TB3
[-> Main menu -> F5: System administration -> F4: System info -> F1: System self test -> F2: System self
test with TB3]
The system self test with TB3 checks the calibration values of the UNIMET® 1100ST. Please
note the following conditions to be met in order for the test to be carried out:
●
Connect the UNIMET® 1100ST to an earthed mains supply (not an IT system).
●
You need the optional TB3 test box. When connecting and using the TB3 test box, please note
the information on the instruction leaflet!
●
Do not connect any earthed printers.
1. Press Enter to start the self test.
2. Check the assignment of the jumpers on the TB3 and then connect the TB3 to the UNIMET®
1100ST as indicated:
3. Press Enter to continue the self test. Follow the instructions that appear on the UNIMET® 1100ST
display.
The test takes approximately one minute to complete. The result of the test is displayed once
the test is complete. You can print out and save the test results for certification purposes.
Test steps during which the limiting value was not observed are indicated by *. If the
UNIMET® 1100ST fails the internal test, contact BENTRON
®
Service.
,
Select master data for display or modification.
Show detailed information about the limiting values.
Modify selected master data.
Exit self test.
i
a
b
c
..
<<
⇒
Содержание UNIMET 1000ST
Страница 50: ...Operation and setting 50 TGH1256en 10 2005...
Страница 80: ...PC compatible functions of the UNIMET 1100ST 80 TGH1256en 10 2005...
Страница 82: ...Maintenance and calibration 82 TGH1256en 10 2005...
Страница 90: ...Options and accessories 90 TGH1256en 10 2005...
Страница 96: ...Data 96 TGH1256en 10 2005...
Страница 99: ......