Options and accessories
89
Commissioning
Prior to first use, the VK701 adapter must be zero balanced. Connect the VK701 measuring
cables to bushes [B] and [C] and plug the VK701 into the UNIMET® 1100ST socket connector.
Bushes 1 and 10 are not used. Then connect the VK701 plug and coupler. Start zero
balancing (See “Test probe, VK701” on page 43).
Please see the VK701 data sheet for more information.
8.6 TP1010 for tests to IEC 61010-1:2001-02
For tests in accordance with standard IEC 61010-1:2001-02, UNIMET® 1100ST always needs
to use the TP1010 test probe. International standard IEC 61010-1:2001-02 sets out general
safety requirements for electrical equipment for measurement, control and laboratory use.
Please see the TP1010 instruction leaflet for more information.