Others
83
No
Item
Frequency Inspection method
and diagnosis
Influence if inspection
not conducted
Maintenance required in
case of nonconformity
9
Care
Vacuum
and saliva
ejector
handpiece
After
closing
Flush the suction line, and
then clean the filter of the
vacuum or saliva ejector
handpiece
Faulty suction may arise
Clean the suction line and
filter in accordance with
"Method for care"
described in the instruction
manual.
10
Care
Cuspidor
section
After
closing
Clean the cuspidor and dust
filter.
Faulty water drainage
may arise.
Execute dust removal and
cleaning in accordance
with "Method for care"
described in the instruction
manual.
11
Care Solid
collector
After
closing
Clean the filter of the solid
collector.
Vacuum suction will
become weak.
Clean the filter in
accordance with "Method
for care" described in the
instruction manual.
12
Care
Exterior
After
closing
Chemical, filthy water and so
forth shall not be found
(attached or remaining) on the
product exterior.
Discoloration and
deterioration to the
exterior, and corrosion
and rusting to metallic
components may arise.
Execute wiping in
accordance with "Method
for care" described in the
instruction manual.
13
Check of
main
switch and
main
valves
After
closing
The product main switch shall
be off, and water/air main
valves shall be closed.
Product failure and
troubles may arise.
Contact your dealer or our
office if the main switch
cannot be turned off or if
the main valve cannot be
closed.
14
Product's
moving
parts
Once every
week
No abnormal noise or the like
shall be produced from
product's moving parts when
the product is operated.
The product will not
normally work and
troubles may arise.
Contact your dealer or our
office if any abnormality
arises.
15
Care Drain
valve
Water may enter the air line,
and equipment failure may
arise.
Drain the water from the
air filter drain valve.
Drain the water from the
air filter in accordance
with "Method for care"
described in the instruction
manual.
16
Check of
Water
pressure
and air
pressure
Check the water pressure and
air pressure by reading
pressure gauges in the
cuspidor section.
Basic set pressure :
Water: 0.1 to 0.2 MPa
Air: 0.45 to 0.5 MPa
The product will not
normally work, and
troubles may arise.
Contact your dealer or our
office if the set pressure is
abnormally high or low.
17
Check of
conditions
of table
section
The table shall be free of
inclination, and water shall not
flow on the table.
Injury caused by falling
of goods located on the
table and other troubles
may arise.
Contact your dealer or our
office if any abnormality
arises.
18
Care Oil
mist
separator
The oil level in the oil mist
separator shall be lower than
the red line.
Normal output will not
be produced due to
inferior handpiece
exhaust.
Discharge the oil in
accordance with "Method
for care" described in the
instruction manual.
MAINTENANCE AND INSPECTION
Содержание CP-ONE PLUS
Страница 12: ...Chair part 8 OVERVIEW AND MAJOR COMPONENTS Seat Legrest Roll up Legrest...
Страница 24: ...Unit part 20 Base mount type Cart type OVERVIEW AND MAJOR COMPONENTS...
Страница 25: ...Unit part 21 OVERVIEW AND MAJOR COMPONENTS...
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ...NOTE BOOK NO 1A07NMME Printed in Japan 2017 09...