Unit part
65
Chair operation stopping function
▪ The safety mechanism for stopping chair operation
is activated in the following cases.
1. When the foot controller pedal is stepped on.
2. When the spittoon bowl is turning toward the
patient (Automatic operation and manual
upward movement stop.)
* In some cases, other display functions may
give information. Be sure to read page 49.
Handpiece rotation stopping function
▪ The handpiece does not rotate if it is removed from
the holder under the condition where you step on
the foot controller pedal. At this time, the message
shown on the left is displayed on the indicator,
showing that the safety mechanism is activated. Be
sure to pick up the handpiece first and then step on
the foot controller pedal.
Numbers and locations to indicate safety
functions
▪ When the safety mechanism is activated, the
position where the safety switch is activated is
displayed on the indicator.
Chair01: Below roll-up
Chair02: Below flange
Chair03: Below main link
Table01: Table raise
Table02: Table rotation
Asst.01: ASST holder rotation
Bowl>>>: Bowl rotation
When any switch in the
dotted line frame is
pressed during automatic
operation, chair operation
stops.
OPERATION STOPPING FUNCTION
DA-290N
Содержание CP-ONE PLUS
Страница 12: ...Chair part 8 OVERVIEW AND MAJOR COMPONENTS Seat Legrest Roll up Legrest...
Страница 24: ...Unit part 20 Base mount type Cart type OVERVIEW AND MAJOR COMPONENTS...
Страница 25: ...Unit part 21 OVERVIEW AND MAJOR COMPONENTS...
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ...NOTE BOOK NO 1A07NMME Printed in Japan 2017 09...