Unit part
60
Setting of the functions exclusive to NLX PLUS
Setting of NLX PLUS torque, automatic reverse, and automatic forward
▪ The
NLX PLUS
micromotor has functions of
setting the torque at the low revolution range (100
to 5000 rpm), activating/deactivating the automatic
reversing and automatic forward.
▪ The functions of setting torque, automatic reverse
and automatic forward are exclusive to the
NLX
PLUS
micromotor.
[Setting procedures]
1.
Remove NLX PLUS from the holder.
2. Press the page feed button
twice and select
.
3. To change the set torque, select “TRQ SET” and
press
to determine.
4. Select intended setting with
or
and
press
to save. The torque is selectable within
the range from 0.3 Ncm (minimum) to 3.0 Ncm
(maximum).
5. To activate/deactivate automatic reverse or to set
the automatic forward, select “AUTO R/F SET”
and press
to determine.
6. Select intended setting with
or
and
press
to save.
▪ Setting is stored in memory (M1 to M4).
CAUTION
Be sure to turn off the automatic reverse and switch the rotation direction using the micromotor
forward/reverse switch
when root canal therapy is performed using NLX PLUS with a left-handed
screw-shaped file. Use with the automatic reverse on may cause fracture of files.
TRQ SET
AUTO R/F SET
TRQ SET
0.3-3.0
0.3
AUTO R/F SET
REVERSE OFF
– REVERSE ON –
FORWARD ON
ON
OPERATION SETTING
N-PLUS
N-PLUS
N-PLUS
Содержание CP-ONE PLUS
Страница 12: ...Chair part 8 OVERVIEW AND MAJOR COMPONENTS Seat Legrest Roll up Legrest...
Страница 24: ...Unit part 20 Base mount type Cart type OVERVIEW AND MAJOR COMPONENTS...
Страница 25: ...Unit part 21 OVERVIEW AND MAJOR COMPONENTS...
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ...NOTE BOOK NO 1A07NMME Printed in Japan 2017 09...