SW 32 Ti and SW 32.1 Ti Swinging-Bucket Rotors
Returning a Rotor
Decontamination
If the rotor or other components are contaminated with toxic, radioactive, or pathogenic materials,
follow appropriate decontamination procedures as outlined by your laboratory safety officer. Refer
to Appendix A in
Rotors and Tubes
to select solutions that will not damage the rotor and accessory
materials.
Sterilization and Disinfection
•
The rotor and all rotor components, except those made of Noryl, can be autoclaved at 121°C for
up to an hour. Remove the plugs from the rotor and place the rotor, plugs, and spacers in the
autoclave upside down.
•
Ethanol (70%)
*
or hydrogen peroxide (6%) may be used on all rotor components, including
those made of plastic. Bleach (sodium hypochlorite) may be used, but may cause discoloration
of anodized surfaces. Use the minimum immersion time for each solution, per laboratory
standards.
While Beckman Coulter has tested these methods and found that they do not damage the rotor or
components, no guarantee of sterility or disinfection is expressed or implied. When sterilization or
disinfection is a concern, consult your laboratory safety officer regarding proper methods to use.
Refer to publication IN-192 (included with each box of tubes) for tube sterilization and disinfection
procedures.
Quick-Seal, Ultra Clear, and thinwall open-top tubes are disposable and should be discarded
after a single use.
Storage
When it is not in use, store the rotor and buckets in a dry environment (not in the instrument).
Remove the bucket caps to allow air circulation so that moisture will not collect in the buckets.
Returning a Rotor
Before returning a rotor or accessory for any reason, prior permission must be obtained from
Beckman Coulter, Inc. This form may be obtained from your local Beckman Coulter sales office. The
form, entitled
Returned Material Authorization
(RMA) for United States returns or
Returned Goods
Authorization
(RGA) for international returns, should contain the following information:
•
rotor type and serial number,
•
history of use (approximate frequency of use),
•
reason for the return,
* Flammability hazard. Do not use in or near operating ultracentrifuges.