LXL-TB-015BB
7
SW 32 Ti and SW 32.1 Ti Swinging-Bucket Rotors
Preparation and Use
Removal and Sample Recovery
CAUTION
If disassembly reveals evidence of leakage, you should assume that some fluid
escaped the rotor. Apply appropriate decontamination procedures to the
centrifuge and accessories.
1
Grasp the bucket cap tabs and lift up to remove the buckets.
•
(Buckets may be removed without removing the rotor body from the ultracentrifuge.)
2
Remove the bucket caps by twisting counterclockwise and lifting up.
a.
Use a tube removal tool (361668) or hemostat to remove the spacers and tubes.
NOTE
If the conical-shaped adapters that support
k
onical tubes are difficult to remove after
centrifugation, an extractor tool (354468) is available to facilitate removal.
3
To remove the rotor from the centrifuge, lift the rotor straight up and off the drive hub.
a.
Set the rotor on the rotor stand.
Tubes and Accessories
The SW 32 Ti rotor uses tubes and accessories listed in
Table 1
; the SW 32.1 Ti rotor uses tubes and
accessories listed in
Table 2
. Be sure to use only those items listed, and to observe the maximum
speed limits shown. Refer to Appendix A in
Rotors and Tubes
for information on the chemical
resistances of tube and accessory materials.
1.
Extractor Tool (354468)
While pressing the rubber tip against
the adapter wall, pull the tube and
adapter up and out of the cavity.