LXL-TB-015BB
11
SW 32 Ti and SW 32.1 Ti Swinging-Bucket Rotors
Preparation and Use
1
Place the tubes in the rack and fill each tube to the base of the stem, leaving no fluid in the stem.
•
Overfilling the tube can cause spillage when the plug is inserted or can compromise seal
integrity.
•
However, too much air can cause excessive tube deformation, disrupting gradients and
sample bands.
2
Refer to
Using OptiSeal™ Tubes
(publication IN-189), included in each box of tubes, for detailed
information on the use and care of OptiSeal tubes.
Quick Seal Tubes
Quick-Seal® tubes must be sealed prior to centrifugation. These tubes are heat sealed and do not
need caps; however, spacers are required on top of the tubes when they are loaded into the rotor
buckets.
1
Fill Quick-Seal tubes leaving a
small
bubble of air at the base of the neck.
a.
Do not leave a large air space—too much air can cause excessive tube deformation.
2
Some of the tubes listed in
Table 1
and
Table 2
are part of the
g
-Max™ system, which uses a
combination of small bell-top Quick-Seal tubes and floating spacers (also called
g
-Max spacers).
•
This means that you can run the shorter tubes listed in the Tables in the SW 32 Ti and
SW 32.1 T rotors without reduction in
g
force.
•
For detailed information on the
g
-Max system see publication DS-709.
1.
Spacer
2.
Plug
3.
Stem
4.
Meniscus
5.
Tube
6.
Base of Stem
1.
g
-Max Spacer
2.
Bell-top Tube
1
5
4
3
2
6
1
2
Содержание SW 32 Ti