Temperature Limits
•
Plastic tubes have been centrifuge tested for use at temperatures between 4 and 25°C. For
centrifugation at other temperatures, pretest tubes under anticipated run conditions.
•
If plastic containers are frozen before use, make sure that they are thawed to at least 4°C prior
to centrifugation.
OptiSeal Tubes
OptiSeal™ tubes come with plastic plugs and can be quickly and easily prepared for use. With the
tube spacer in place, the
g
force during centrifugation ensures a tight, reliable seal that protects
your samples.
16
67 mm
10.5 mL
polyallomer
Quick-Seal, bell-top
344622
(pkg/50)
Noryl floating spacer
355579
32,000 RPM
187,000
g
133
16
57 mm
8 mL
polyallomer
Quick-Seal, bell-top
344621
(pkg/50)
Noryl floating spacer
355579
32,000 RPM
187,000
g
102
16
44 mm
6.5 mL
polyallomer
Quick-Seal,
bell-top
345830
(pkg/50)
Noryl floating spacer
355579
32,000 RPM
187,000
g
79
16
32 mm
4.5 mL
polyallomer
Quick-Seal,
bell-top
356562
(pkg/50)
Noryl floating spacer
355579
32,000 RPM
187,000
g
56
a. Use only the items listed here..
Table 2
Beckman Coulter Tubes and Accessories for the SW 32.1 Ti Rotor
a
(Continued)
Tube
Required Accessory
Max Speed/
RCF/
k
factor
Dimensions/
Nominal
Volume/
Description
Part
Number
Description
Part
Number
4°C