3 Disassembly / Assembly
Infusomat® compact
plus
P 1.0
3 - 11
EN
For internal use, only
3.7
PUMP DOOR
Designation Ord.
No.
Pump door ICPP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34522294
Screws
Disassembly
✓
Pump door is open.
1. Unscrew the three screws (
) using a 15IP
TORX plus screwdriver.
2. Lift the hinges (
) from the housing, bottom
part (
3. Remove the pump door (
).
Assembly
✓
Unit is closed.
✓
Second clamp has been mounted.
1. Tighten the screws with a torque of 0.7 Nm ± 0.07 Nm (first
and second screwing).
Fig. 3 - 11
1 Pump door
2 Hinge
3 Screw DELTA PT 35x10 WN 5452
4 Housing, bottom part
NOTICE
The complete check after repair has to be performed whenever
the pump door has been removed.
Содержание 8717070
Страница 1: ...Infusomat compactplus P Service Manual EN Version 1 0 For internal use only For internal use only...
Страница 14: ...Important Preliminary Remarks 0 14 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 16: ...Contact Information 0 16 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 24: ...1 Overview of the Device 1 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 32: ...2 Unit Diagnosis Adjustments 2 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 80: ...5 Servicing the Unit 5 2 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 84: ...6 Technical Safety Check TSC 6 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 94: ...7 Procedural Instructions on the TSC 7 10 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 100: ...9 Spare Parts List 9 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 105: ......