4 Check after Repair
Infusomat® compact
plus
P 1.0
4 - 7
EN
For internal use, only
T2 Electrical safety according to IEC 62353
The values to be measured for electrical safety are listed in the
checklist
(see "Checklist for Check after Repair"
Equipment leakage current (alternative method):
1. Measure between mains input and pin 3 (GND) of accessory
connector (
).
For safer measurement, use the Measuring Adapter Accessory
Connector CP.
T3 Mechanical inspection
1. Fit the unit to be tested on top of another compact
plus
device
and check that the lock functions correctly.
2. Fit the unit to be tested underneath another compact
plus
device
and check that the lock functions correctly.
3. Rotate the pole clamp in all directions (4x90°) and check for
correct latching.
4. Check that the rotary knob of the pole clamp can be turned
clockwise and counterclockwise.
T4 Operational test
1. Connect the unit to the power supply without switching it on
and check the status display:
a) Battery capacity
b) Mains operation symbol
2. Switch the unit on and check the following details:
a) Self-test
b) Display of software version
c) Audible alarm
one high sound from piezo buzzer
one deep sound from loudspeaker
d) Visual alarm
The green LED flashes once. The red LED flashes twice.
Fig. 4 - 1
1 Pin 3 of accessory connector
▲
Top of unit
▼
Bottom of unit
Содержание 8717070
Страница 1: ...Infusomat compactplus P Service Manual EN Version 1 0 For internal use only For internal use only...
Страница 14: ...Important Preliminary Remarks 0 14 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 16: ...Contact Information 0 16 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 24: ...1 Overview of the Device 1 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 32: ...2 Unit Diagnosis Adjustments 2 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 80: ...5 Servicing the Unit 5 2 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 84: ...6 Technical Safety Check TSC 6 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 94: ...7 Procedural Instructions on the TSC 7 10 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 100: ...9 Spare Parts List 9 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 105: ......