4 Check after Repair
4 - 2
Infusomat® compact
plus
P 1.0
EN
For internal use, only
PUMP RUN IN
Performing the RUN IN
1. To run-in the drive, insert an Intrafix Primeline
Do not close the pump door, yet.
2. Use the "Menu" button and the arrow keys to select the
"Settings" menu item and confirm with "OK".
3. Select "Service" and confirm the device code with "OK".
4. Confirm selection of "All Disposables Off" with "OK" (
5. Select "On" with "OK".
"Service active" is displayed.
6. Confirm " Do not use at Patient" with "OK" (
7. Close the pump door.
NOTICE
The "All Disposables On" function has to be activated before the
pump door is closed.
Fig. 4 - 1
Fig. 4 - 2
Содержание 8717070
Страница 1: ...Infusomat compactplus P Service Manual EN Version 1 0 For internal use only For internal use only...
Страница 14: ...Important Preliminary Remarks 0 14 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 16: ...Contact Information 0 16 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 24: ...1 Overview of the Device 1 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 32: ...2 Unit Diagnosis Adjustments 2 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 80: ...5 Servicing the Unit 5 2 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 84: ...6 Technical Safety Check TSC 6 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 94: ...7 Procedural Instructions on the TSC 7 10 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 100: ...9 Spare Parts List 9 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 105: ......