(Master – to be added to the documentation)
4
Check after Repair
Infusomat® compact
plus
P 1.0
4 - 5
EN
For internal use, only
YES N/A
T7 SHK
❒ ❒
❒
Is closed after opening the pump door.
❒
P
SHK
(> 0.5 bar)
_____ bar
(or > 7.3 psi)
_____ psi
YES N/A
T8 Second clamp
❒ ❒
❒
Is closed after opening the pump door.
❒
P
2ndClamp
(> 0.5 bar)
_____ bar
(or > 7.3 psi)
_____ psi
YES N/A
T9 Delivery accuracy
❒ ❒
Ambient temperature 20 … 25°C,
delivery rate: 250 ml/h
❒
Divergence (
±
5 %)
_____ %
YES N/A
T10 Air sensor
❒ ❒
❒
Water value
(1100 … 2250 mV)
_____ mV
❒
Air alarm
❒
Air value (< 100 mV)
_____ mV
YES N/A
T11 Temperature sensor
❒ ❒
❒
FuP Temp - KuP Temp
(-1.0 … +1.0 °C)
_____ °C
YES N/A
T12 Staff Call Test
❒ ❒
Connect the staff call cable compact
plus
.
Start infusion and open the pump door.
❒
An alarm goes off
❒
Red LED of staff call cable compact
plus
lights up
YES N/A
Fixing the Seal
❒ ❒
❒
Pump is sealed with a new seal
Mech. Aids and Measuring Equip. Used
Accessories used
❒
Disposables
Type: ________________________
Part No.: _____________________
❒
Manometer
Ident. No.: ____________________
Calibrated until: _______________
❒
Scales
Ident. No.: ____________________
Calibrated until: _______________
❒
Safety tester
Ident. No.: _____________________
Calibrated until: ________________
❒
Measuring Adapter Accessory
Connector CP
❒
Graduated cylinder
❒
Stop watch
❒
___________________________
❒
___________________________
❒
Mains power supply cable
❒
Staff call cable compact
plus
❒
Connection lead 12V
❒
Infusion line
❒
__________________________
❒
__________________________
Test result:
Defects found that could endanger patients, users, or
third parties:
Yes No
Check performed by:
Measures to be taken:
None
Unit handed over on:
Repair
________________________________
To:
Special features / documentation:
___________________________________________________________
Date / Signature:
(Sheet 2 of 2)
Содержание 8717070
Страница 1: ...Infusomat compactplus P Service Manual EN Version 1 0 For internal use only For internal use only...
Страница 14: ...Important Preliminary Remarks 0 14 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 16: ...Contact Information 0 16 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 24: ...1 Overview of the Device 1 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 32: ...2 Unit Diagnosis Adjustments 2 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 80: ...5 Servicing the Unit 5 2 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 84: ...6 Technical Safety Check TSC 6 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 94: ...7 Procedural Instructions on the TSC 7 10 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 100: ...9 Spare Parts List 9 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 105: ......