3 Disassembly / Assembly
Infusomat® compact
plus
P 1.0
3 - 5
EN
For internal use, only
3.2
GASKET FOR RUBBER CONNECTOR
Designation Ord.
No.
Gasket for rubber connector CP. . . . . . . . . . . . . . . . . . . . .34522010
Screws
Disassembly
1. Unscrew the four screws (
) using a 10IP TORX
plus screwdriver.
2. Remove the gasket (
).
Assembly
✓
Power supply is installed in the housing, bottom part.
1. Attach the gasket with the correct orientation.
2. Tighten the screws with a torque of 0.75 Nm ± 0.1 Nm.
Fig. 3 - 3
1 Screw DELTA PT 30x9 WN 5454
2 Gasket
NOTICE
Use a new gasket whenever the gasket has been removed.
Содержание 8717070
Страница 1: ...Infusomat compactplus P Service Manual EN Version 1 0 For internal use only For internal use only...
Страница 14: ...Important Preliminary Remarks 0 14 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 16: ...Contact Information 0 16 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 24: ...1 Overview of the Device 1 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 32: ...2 Unit Diagnosis Adjustments 2 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 80: ...5 Servicing the Unit 5 2 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 84: ...6 Technical Safety Check TSC 6 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 94: ...7 Procedural Instructions on the TSC 7 10 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 100: ...9 Spare Parts List 9 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 105: ......