3 Disassembly / Assembly
3 - 10
Infusomat® compact
plus
P 1.0
EN
For internal use, only
3.6
SECOND CLAMP
Designation Ord.
No.
Second clamp ICPP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34522286
Screw and cover cap
Disassembly
✓
Pump door is open.
1. Pierce the cover cap (
) with a small screw-
driver or twist gimlet and remove it.
) using a 10IP TORX plus
screwdriver.
3. Slightly lift the second clamp (
) toward the
top and remove it to the side.
Assembly
✓
Unit is closed.
1. Tighten the screw with a torque of 0.65 Nm ± 0.1 Nm (first and
second screwing).
Fig. 3 - 10
1 Cover cap
2 Screw DELTA PT 30x10 WN 5452
3 Second clamp
NOTICE
If the second clamp ICPP has to be replaced, the transport
bracket must be removed from the new spare part before it is
mounted to the pump.
Содержание 8717070
Страница 1: ...Infusomat compactplus P Service Manual EN Version 1 0 For internal use only For internal use only...
Страница 14: ...Important Preliminary Remarks 0 14 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 16: ...Contact Information 0 16 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 24: ...1 Overview of the Device 1 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 32: ...2 Unit Diagnosis Adjustments 2 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 80: ...5 Servicing the Unit 5 2 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 84: ...6 Technical Safety Check TSC 6 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 94: ...7 Procedural Instructions on the TSC 7 10 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 100: ...9 Spare Parts List 9 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 105: ......