Important Preliminary Remarks
Infusomat® compact
plus
P 1.0
0 - 9
EN
For internal use, only
MEASURING EQUIPMENT
Additional measuring and test equipment used for servicing our
devices must comply with the following standards and/or mea-
suring equipment specifications:
Electrical safety:
IEC 60601-1, ed 3.1
EN 62353
For example, Bender UNIMET 800/810ST
bender-de.com/en
or comparable
Voltage measurement: For example, Multimeter Fluke 177
fluke.com/en-us
or comparable
Clearance measurement:
Slip gauges
According to DIN 2275, tolerance T2 or better.
For example,
wermas.de
or comparable
Pressure measurements:
Measurement accuracy
≤
80 mbar (1.16 psi)
For example, Kobold MAN-SD1S Y A3 0
S#2062378.8 customer-specific
kobold.com/en
or Sika MH 3750
sika.net
or Fluke 717
fluke.com
or comparable
Flow accuracy:
Scales: Measurement
range
220 g
Linearity deviation
±
0.1 mg
Standard deviation
±
0.05 mg
Readability 0.01 mg
Содержание 8717070
Страница 1: ...Infusomat compactplus P Service Manual EN Version 1 0 For internal use only For internal use only...
Страница 14: ...Important Preliminary Remarks 0 14 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 16: ...Contact Information 0 16 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 24: ...1 Overview of the Device 1 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 32: ...2 Unit Diagnosis Adjustments 2 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 80: ...5 Servicing the Unit 5 2 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 84: ...6 Technical Safety Check TSC 6 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 94: ...7 Procedural Instructions on the TSC 7 10 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 100: ...9 Spare Parts List 9 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 105: ......