(Master – to be added to the documentation)
EN
4 - 4
Infusomat® compact
plus
P 1.0
For internal use, only
4
Check after Repair
Checklist for Check after Repair
Checklist for Check after Repair – After Repair or Calibration
Owner
Unit: Infusomat® compact
plus
P
Manufacturer: B. Braun Melsungen AG
Year of Procurement
Observe the Service Manual and the Instructions for Use. All measured values are
to be documented. Accessories used should be included in testing. Only use calibrated
measuring equipment.
Software Version
Article No. (REF)
Serial No. (SN)
Inventory No. (of the Owner)
YES N/A
T1 Visual Inspection
❒ ❒
❒
Infusomat® compact
plus
P:
Cleanliness, completeness, damage,
and defects affecting safety, damage
and readability of the labels.
Particularly:
❒
Housing
❒
Screw covers
❒
Pump door
❒
SHK ICPP
❒
Air sensor (free from grease or
ultrasonic gel, free from cracks)
❒
Pressure sensor, downstream
(membrane)
❒
Pump membrane
❒
Pressure sensor, upstream
(membrane)
❒
Second clamp
❒
Keypad/display
❒
Accessory connector
❒
IR window
❒
Mains power supply connection
❒
Accessories
Cleanliness, completeness, damage,
and faults affecting safety, damage,
and readability of the labels
❒
Check the unit and the accessories for
compatibility
❒
Mains power supply cable
YES N/A
T2 Electrical Safety
❒ ❒
❒
Protective conductor resistance of
mains connecting line
(
≤
0.1
Ω)
______
Ω
❒
Measure
mains voltage
_____ V~
❒
Equipment leakage current
(alternative method for devices of
protection class II)
(≤
10
μ
A)
_____
μ
A
Measure between mains input and
pin 3 of accessory connector. Use the
Measuring Adapter Accessory
Connector CP.
YES N/A
T3 Mechanical Inspection
❒ ❒
❒
Locking with second unit
❒
Pole clamp
YES N/A
T4 Operational Test
❒ ❒
Connect unit to mains power supply
❒
Status display
Switch on unit with mains power supply
❒
Self-test
❒
Information on display
❒
Audible alarm
❒
Visual alarm
YES N/A
T4 Operational Test (cont.)
❒ ❒
Switch on unit without mains power supply
❒
Self-test
❒
Battery test
❒
Opening and closing mechanism of the
operating unit
❒
Infusion
❒
Buttons on the operating unit
❒
Trigger bolus at the device
❒
Staff call (if available)
YES N/A
T5 Pressure cut-off, electronic
❒ ❒
Delivery rate: 250 ml/h.
❒
Upstream
❒
Alarm
❒
Downstream
❒
Pressure stage 5
(0.1 … 0.7 bar)
_____ bar
(or 1.5 … 10 psi)
_____ psi
❒
Pressure stage 9
(0.7 … 1.3 bar)
_____ bar
(or 10 … 19 psi)
_____ psi
YES N/A
T6 Pressure limitation, mechanical
❒ ❒
Delivery rate: 250 ml/h.
❒
P
max
(2.4 … 4.0 bar)
_____ bar
(or 35 … 58 psi)
_____ psi
(Sheet 1 of 2)
Edit
Содержание 8717070
Страница 1: ...Infusomat compactplus P Service Manual EN Version 1 0 For internal use only For internal use only...
Страница 14: ...Important Preliminary Remarks 0 14 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 16: ...Contact Information 0 16 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 24: ...1 Overview of the Device 1 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 32: ...2 Unit Diagnosis Adjustments 2 8 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 80: ...5 Servicing the Unit 5 2 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 84: ...6 Technical Safety Check TSC 6 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 94: ...7 Procedural Instructions on the TSC 7 10 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 100: ...9 Spare Parts List 9 4 Infusomat compactplus P1 0 EN For internal use only For your notes...
Страница 105: ......