CPI.2
Revised 5/01
Pin Assignments for Interface Cable — Serial Port
Pin
Signal
Signal
Number
Name
Function
Type
Source
1
CHAS GND
Protective Ground
Ground
N/A
2
TXD
Transmit Data
Data
CLINITEK 500
3
RXD
Receive Data
Data
Computer
4
RTS
Request To Send
Control
CLINITEK 500
5
CTS
Clear To Send
Control
Computer
6
DSR
Data Set Ready
Control
Computer
7
SIG GND
Signal Ground
Ground
N/A
20
DTR
Data Terminal Ready
Control
CLINITEK 500
(All other pins are unused.)
Hardware Handshaking:
RXD
Receive Data
This input receives control characters for software handshaking and data for IDs.
TXD
Transmit Data
This output sends test data, control characters, and instrument information.
DTR
Data Terminal Ready
This signal is on whenever the instrument I/O is configured for a computer and the
instrument is
on.
DSR
Data Set Ready
The computer must raise this line whenever it is ready to receive data. If not supplied
by the computer, pin 6 must be jumpered to pin 20.
RTS
Request To Send
This output line, when high, indicates to the computer that it may send data.
CTS
Clear To Send
This input is checked before sending each character and, if high, the next character
is sent. If not supplied by the computer, pin 5 must be jumpered to pin 4.
The following signal lines are NOT implemented:
DCD
Data Carrier Detect
Pin 8
RNG
Ring Indicator
Pin 22
Table CPI-1
COMPUTER/PRINTER INTERFACE
40069/CPI 11/8/06 10:00 AM Page 2
Содержание Clinitek 500
Страница 1: ...Urine Chemistry Analyzer Operating Manual 63032 Cover 11 8 06 9 10 AM Page 1...
Страница 22: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 40: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 60: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 72: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 76: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 82: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...