4.2
Revised 5/01
Before starting each run,
perform the following steps:
1. Check that the name of the Reagent Strip being dis-
played corresponds to the name of the Bayer Reagent
Strip for Urinalysis that is to be used for routine urin-
alysis.
Use of any Reagent Strip other than the one
listed will cause erroneous results.
If the names do
not agree, select the correct strip name through the
Setup
Routine, as directed in Section 3, Step C-2.
1.
IMPORTANT:
The CLINITEK 500 Analyzer has been
optimized for use with Bayer Brand Reagent Strips.
The performance characteristics may not be valid for
any other brand of strips, the use of which may cause
erroneous results. Bayer Corporation does not war-
rant use of the CLINITEK 500 Analyzer with any
reagent strip other than Bayer Brand Reagent Strips.
2. Inspect the strip loading station and push bar for
cleanliness and correct positioning. If contaminants
are present, remove the push bar, platform, and mov-
ing table, and clean as described in Section 5, “Daily
Cleaning.”
3. The “SEQ #” (sequence number) increments with each
strip that is placed onto the instrument. Change the
starting number if desired:
3.
a. Touch the key symbol next to the word
SEQ #.
The
display will change to a numeric keyboard, from
which the new number can be entered.
3.
b. Enter the desired number. The number can be reset
to “00001” by touching the special key labeled with
that number. If an error is made, or if you only need
to change one or two digits, touch the appropriate
key (
or
) to move the cursor to the digit to
be changed and enter the correct number.
3.
c. When the correct number has been entered, touch
to enter the number into memory.
4. Change the “Tech ID” (technician identification) if
needed:
4.
NOTE:
If the “Tech ID” key is dimmed, the option can
be changed to ON as directed in Section 3, Step H-3.
4.
a. Touch the
Menu
key symbol to display the Option
Menu; for example:
4.
b. Touch the
Tech ID
key. A numeric keyboard will be
displayed.
4.
c. Enter the identification of up to 13 digits. The key-
board also includes the
key, which allows
alphabetic characters to be entered. Touch
from the alphabetic screen to return to the
numeric keyboard, then touch
to store the Tech
ID and return to the Option Menu screen.
5. If you want to print the ID list (if a loadlist exists in
memory), confirmation of the last calibration, or a
report of the setup parameters, touch the
key
from the Option Menu and select the desired option.
Touch
to return to the Option Menu.
6. If you want to run controls before starting the run,
touch the
Controls
key from the Option Menu and pro-
ceed as directed next.
Set options.
Tech ID: 118
Controls
Setup
Memory
INSTRUMENT OPERATION
63032-Section 4 11/8/06 9:25 AM Page 2
Содержание Clinitek 500
Страница 1: ...Urine Chemistry Analyzer Operating Manual 63032 Cover 11 8 06 9 10 AM Page 1...
Страница 22: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 40: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 60: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 72: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 76: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 82: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...