3.16
Revised 5/01
SELECTING YOUR OPTIONS
2.
e.
Display:
The display can be tested to ensure that
all pixels (the lighted elements on the display) are
being lit and turned off appropriately. When the
Display
key is touched, the entire screen will be lit
for several seconds, then be blank. This series will
be repeated twice more before returning to the
previous menu. If there are numerous faulty pixels,
or if they are located in critical areas, the display
may need to be replaced.
2.
f.
Printer:
The internal and/or 80-column external
printer can be tested to ensure that all characters
are being printed correctly. Ensure that the exter-
nal printer is turned
on (if one is being used). Touch
the
Printer
key and follow the directions on the
screen. Examine the printout for its readability. The
display automatically returns to the previous menu.
3. When all tests have been completed, touch
to return
to any previous
Setup
menu, or touch
to return to
the
Ready/Run
screen.
When Setup is Complete
Touch to
save!
When you have finished selecting the setup parame-
ters, touch
to return the display to the
Ready/Run
screen. The changes you made will NOT be saved if you
do not touch
.
It is suggested that you print a copy of the setup report
to verify your selections and to retain in your files. Touch
Menu
from the
Ready/Run
screen, then touch
Print.
From
the Print menu that is displayed, select the option
Setup
report.
(If printing from the internal printer, make a
photocopy of the report, since the thermal print may fade
over time.)
The setup parameters are stored both in the CLINITEK
500 Analyzer and in the program card. If a new program
card is installed, you can select to use your current setup
configuration from the Analyzer memory. Conversely, if
you receive a new CLINITEK 500 Analyzer, but have
retained your old program card, you can select to use your
current setup configuration from the program card. If both
the Analyzer and the program card are new, you can use
the printout obtained previously to reselect the desired
parameters.
63032-Section 3 11/8/06 9:21 AM Page 16
Содержание Clinitek 500
Страница 1: ...Urine Chemistry Analyzer Operating Manual 63032 Cover 11 8 06 9 10 AM Page 1...
Страница 22: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 40: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 60: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 72: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 76: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 82: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...