8.4
Revised 9/99
TROUBLESHOOTING/SERVICE
TROUBLESHOOTING CHART
SYMPTOM
POSSIBLE CAUSE
REMEDY
Changes made to
The
Return to Ready/Run
key (
)
Always touch
after all changes have been made to
Setup
Routine are
was not touched after changes
the
Setup
Routine.
not saved.
were made.
Display is blank.
1. No power.
1. Listen for the fan; if it is not running, turn the instru-
ment power
off. Check that the power cord is firmly
plugged into the instrument and into a live AC elec-
trical outlet. Then turn the power back
on.
2. Improperly inserted program card.
2. Turn the instrument power
off, then remove the pro-
gram card and reinsert it firmly, making sure that
the label is facing forward, with the arrows pointing
in and up. When properly inserted, the edge of the
card will be flush with the instrument case. Turn the
power back
on.
3. Defective program card or defective
3. Contact your Bayer Customer Service office.
3.
system electronics.
Fixed platform cannot
The moving table is not in the lowest
Turn the power
on and let the system initialize. Ignore
be installed.
position (see NOTE in “General
the error that is displayed. Turn the power
off again,
Information” earlier in this section).
install the platform, then turn the power
on. If still
unable to install the platform, contact your Bayer
Customer Service office.
Printout does not
“Missing” reports have been flagged
The list of flagged reports will be displayed when the
contain all reports.
for the Confirmatory Report, and
run is complete, and the reports will be printed after
“Edit flagged results” is ON.
the End-of-Run Report screens have been exited.
Push bar does not
1. Other strips are being moved along
1. Allow up to 7 seconds to elapse prior to movement
move to the right after
1.
the platform.
1.
of the push bar. The time lapse depends upon
a strip is placed onto
1.
1.
the timing cycle for movement of the strips
the platform.
1.
across the platform.
2. Strip sensor problem.
2. From the
Ready/Run
screen (and with the run com-
2.
pleted), turn the instrument power
off, wait several
seconds, then turn it back
on. If the problem contin-
ues, contact your Bayer Customer Service office.
63032-Section 8 11/8/06 9:51 AM Page 4
Содержание Clinitek 500
Страница 1: ...Urine Chemistry Analyzer Operating Manual 63032 Cover 11 8 06 9 10 AM Page 1...
Страница 22: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 40: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 60: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 72: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 76: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 82: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...