Revised 10/06
8.9
ERROR DISPLAY
POSSIBLE CAUSE
REMEDY
Error 50
Printer Error
Check that your external printer is turned
on and is on-
line. Verify that both ends of the interface cable are
securely connected and check that your printer has
paper.
Error 51
Control results memory (51) or sample
Nearly 200 control result sets or nearly 500 patient
Error 52
results memory (52) is almost full.
result sets have been stored in memory without being
transferred to a computer. Check that your computer is
turned
on, that the interface cable is securely connected
at both ends, and that the setup parameters for the
computer interface are correct. Transfer at least some
of the records. If unable to transfer records, contact
your Bayer Customer Service office.
Error 53
Control results memory (53) or
Two hundred control result sets or 500 patient result
Error 54
sample results memory (54) is
sets have been stored in memory without being
completely full.
sent to a computer. No additional testing can occur
until at least some of the records are transmitted. Check
that your computer is turned
on, that the interface cable
is securely connected at both ends, and that the setup
parameters for the computer interface are correct.
Error 55
Both areas of instrument memory
Print a Setup report (see Section 3, “When Setup is
where the Setup parameters are stored
Complete”) to view the default parameters. If you
have been corrupted. The manufacturer’s
previously printed and saved a copy of the Setup report
defaults have been restored.
of your customized selections, compare the two
reports. Reselect the options that need to be changed.
Error 56
System error.
Turn the instrument power
off, wait several seconds,
then turn it back
on. If the error repeats, contact your
Bayer Customer Service office.
63032-Section 8 11/8/06 9:51 AM Page 9
Содержание Clinitek 500
Страница 1: ...Urine Chemistry Analyzer Operating Manual 63032 Cover 11 8 06 9 10 AM Page 1...
Страница 22: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 40: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 60: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 72: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 76: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...
Страница 82: ...63032 Mfg Warranty 11 8 06 10 13 AM Page 2...