Start of operation
204
Measuring System MAK TIGER 3003 Service manual, SA 131128, 10.03.2020
Correction factor settings:
Setting range = 0 to 100
0 = curve is still proportional, i.e. as before, the level in the pipe is linear to
the FLM signal (see graph)
100 = curve is now 1/X, i.e. the level in the pipe is not linear (1/X function)
to the FLM signal (see graph)
GUIDE to calculating the correct settings:
All the correction factor settings need to be calculated with milk.
It is advisable to start with 20 (for low and medium) with the small intake
quantities. After analyzing the fat levels, a higher or lower setting can then
be selected based on the tendency, i.e. the correct settings need to be cal-
culated in tests on the actual vehicle in any given case.
Graph illustrating the correction factors:
L
e
v
e
l
i
n
p
i
p
e
Proportional
(+ fat)
Crr.factor = 0
FLM Signal
(2.3 V)
Level in
pipe
100 %
1/x
(-fat)
Crr.factor = 100