background image

Device for analysing 

various ingredients

of beer

Article - No : 3572

FermentoFlash

User manual and reference guide

 Date : August 2010

-2-

Contents

Initial Operation 

 

 

1.

Initial Operation / Unpacking and checking package contents ---------

 3

2

Initial Operation / Installation -----------------------------------------------------

 3

3.

Initial Operation / Connect machine to the power  ---------------------------  3

4.

Initial Operation / Connect mains cable  ---------------------------------------

 3

5.

Initial Operation / Connect hose   -----------------------------------------------

 4

6.

Initial Operation / Connect printer  ----------------------------------------------

 4  

7.

Initial Operation / Switch on machine  -----------------------------------------

 4

8.

Initial Operation / Switch on printer, insert roll of paper  ------------------

 4

9.

Initial Operation / Machine operation-------------------------------------------

 4

10.

Initial Operation / Turning the device on / off --------------------------------

 5

11.

Initial Operation / Zero calibration   ---------------------------------------------

 5

Operation----------------------------------------------------------------------------------------

 6

1. 

Operation / Measuring-------------------------------------------------------------

 6

2.

Operation / Rinsing   ---------------------------------------------------------------

 7

3.

Operation / Product Selection / Example    -----------------------------------

 7 

4.

Operation / Cleaning       ----------------------------------------------------------

 7

5.

Operation Setup    ------------------------------------------------------------------

 8

5.1. Return   -------------------------------------------------------------------------

 8

5.2. System Status    --------------------------------------------------------------

 8

5.3. Intensive Cleaning   ----------------------------------------------------------

 8

5.4. Configure Parameters -------------------------------------------------------

 8

5.5. Language Selection  ---------------------------------------------------------

 9

5.6. Adjust pump  ------------------------------------------------------------------

 9

5.7. Measurements per sample  ------------------------------------------------

 9

5.8. Date  ----------------------------------------------------------------------------

 8

5.9. Time  ----------------------------------------------------------------------------

 8

Maintenance

 9

Calibration

 9

6.

Operation / Calibration  -----------------------------------------------------------

 9

6.1. Performing Calibration  -----------------------------------------------------

11

6.2. Return  -------------------------------------------------------------------------

11

6.3. Name   -------------------------------------------------------------------------

12

6.4. Matrix  --------------------------------------------------------------------------

12

6.5. Product Selection  -----------------------------------------------------------

12

6.6. Fast / Precise  -----------------------------------------------------------------

12

6.7. Selection of the Substance to be Calibrated  --------------------------

12

6.8. Entering the Reference Values  ------------------------------------------

12

6.9. Starting B Calibration  ------------------------------------------------------

12

How the FermentoFlash works  ---------------------------------------------------------

13

1.

Structure of the Device  ----------------------------------------------------------

13

2.

Description  -l------------------------------------------------------------------------

14

3.

Sample suction tube  --------------------------------------------------------------

14

4.

Pump  ---------------------------------------------------------------------------------

14

5.

Replacement of the pump head  -----------------------------------------------

14

Содержание FermentoFlash 3572

Страница 1: ...n off 5 11 Initial Operation Zero calibration 5 Operation 6 1 Operation Measuring 6 2 Operation Rinsing 7 3 Operation Product Selection Example 7 4 Operation Cleaning 7 5 Operation Setup 8 5 1 Return 8 5 2 System Status 8 5 3 Intensive Cleaning 8 5 4 Configure Parameters 8 5 5 Language Selection 9 5 6 Adjust pump 9 5 7 Measurements per sample 9 5 8 Date 8 5 9 Time 8 Maintenance 9 Calibration 9 6 O...

Страница 2: ...e power Power connection 230 V 180 VA or 115 V 180 VA The power connection is located on the back side of the device together with the on off switch voltage selector switch and mains fuse First select the correct mains voltage using the voltage selector switch 230 V or 115 V 4 Initial Operation Connect mains cable Ensure that the machine is connected to an earthed socket which complies with safety...

Страница 3: ...tem c Press the Enter button zero calibration starts Progress is shown by means of a moving bar After zero calibration the device is ready for operation 6 Zero calibration is used to calibrate the hardware It is not calibration of the substances in the beer This is performed with the B calibration If you would like to check the zero calibration process please carry out a control measurement with d...

Страница 4: ... the device should be rinsed 3 4 times 5 Operation Setup 8 5 1 Return After the Enter button has been pressed the display returns to the main menu 5 2 System Status After the Enter button has been pressed the following image appears The system status can be printed out if necessary for error diagnosis 5 3 Intensive Cleaning Approx 1 hour Performing Intensive Cleaning a Place the sample suction hos...

Страница 5: ...igh alcohol content e g strong beer In the case of a two point calibration the A calibration is performed with a beer having low substance content in all parameters A Beer and the B calibration is carried out with a high substance content reference beer B beer The A beer is obtained from the B beer by dilution e g 20 distilled water 80 reference beer It is important that this dilution is made usin...

Страница 6: ...Note the capital letters are followed by the small letters e Continue the procedure until the entire word has been entered f Press the Enter button This confirms and stores what has been entered 6 4 Matrix There are four different product matrices available Beer Strong Beer Non Alcoholic Beer etc Please select the corresponding matrix 6 5 Product Selection See 3 Operation Product Selection Example...

Страница 7: ...s they have to be rinsed very thoroughly after use Never operate the FermentoFlash without rinsing or cleaning after finishing the measurements The device contains two sensitive measurement chambers which are kept at a certain temperature If beer precipitates in such a measurement chamber and is deposited on the walls the measurement chamber will be destroyed 14 How the FermentoFlash works 1 Struc...

Страница 8: ...ing gas bubbles Remove the CO2 content using appropriate measures Procedure Manual Shaking swirling folded filter Electric Mixer folded filter Ultrasonic bath approx 3 minute Attention keep the bath temperature cool with ice cubes 2 The sample must be liquid It must not contain any solids 3 Sample temperature From 5 C to 35 C The beer samples should preferably show a standard temperature The Ferme...

Страница 9: ...ated 18 Sample flow rate up to 100 h typical 60 h Sample volumes 10 ml Interfaces 1 x parallel 1 x serial RS 232 9600 Baud 6 Volt power supply for thermal printer Item No 7151 Dimensions 30 x 24 x 33 cm W x H x D Weight approx 5 kg net Ordering data 3572 FermentoFlash 3563 Cleaner 0 5 litre 8818 Erlemeyer flask 100 ml 7151 Thermal printer incl 1 roll thermal paper items marked with are included in...

Страница 10: ...ty requirements of the EC Machines Directive in its design and construction and in the configuration put on the market by us This declaration is invalid if the machine is modified without our agreement Machine description FermentoFlash Machine type Beer analyser Machine No 3572 Applicable EC Directives EC Machines Directive 06 42 EC EC Low Voltage Directive 73 23 EC EC Electromagnetic Compatibilit...

Отзывы: