Installation position:
Horizontal/flat, rising slightly in the direction of flow so as to allow trapped
gases to escape, or vertical.
May only be operated with flow-calming section on both sides of L
2 x D.
There must be not be any mechanical forces on the flow rate meter either
during assembly/disassembly or during operation. The pipes which are con-
nected to the unit on both sides must therefore be duly supported (see dia-
gram below).
Installation location:
At the lowest point of the pipe system, rising slightly in the direction of flow
Position at which the MID is always filled with liquid.
Reduce amount of gas in the liquid to a minimum as entrained air can also
affect the flow measurement. Therefore install the flow rate meter in the
pipe system at the point of maximum pressure where the air volume is
minimal.
After air separator, if air can be sucked in.
The ambient temperature at the place of installation must be within the
permissible range of -10...+50°C. The liquid flow (measuring medium,
cleaning fluid) through the device must not exceed the maximum
temperature for CIP cleaning 100 ° C.
Position the flow rate meter so that it can be easily accessed for servicing.
The electrodes in the MID always have to be aligned vertically.
Weld the pipeline support,
for example by means of
clamps with bracket to
cabin wall / floor
MID
Pipeline
Содержание MAK TIGER 3003
Страница 13: ...Safety precautions 5 Measuring System MAK TIGER 3003 Service manual SA 131128 10 03 2020...
Страница 31: ...Measuring system MAK TIGER 3003 23 Measuring System MAK TIGER 3003 Service manual SA 131128 10 03 2020...
Страница 32: ...Measuring system MAK TIGER 3003 24 Measuring System MAK TIGER 3003 Service manual SA 131128 10 03 2020...
Страница 187: ...ULTRASAMPLER sample collection system 179 Measuring System MAK TIGER 3003 Service manual SA 131128 10 03 2020 Dimensions...
Страница 216: ...Measuring System MAK TIGER 3003 Service manual SA 131128 10 03 2020...