ULTRASAMPLER® sample collection system
181
Measuring System MAK TIGER 3003 Service manual, SA 131128, 10.03.2020
26.8.2
Wiring
Proximity switch
NPN Order.No. U495593
Ground switching
Controller 1)
MAK compact-Con-
troller
6942-10
US-Controller
6771-31
brown (bn) + 24V_S
IN 1
– 8
+ 24 V clamp 25
blue (bu) AGND
IN 1 - 8
GND clamp 34
black (bk)
Free inputs IN 1 - 8
Free inputs
IN 1
– 8
clamp 26 ... 33
Proximity switch
PNP Order.No. U496004
Positiv switching
Controller 1)
MAK Compact-Con-
troller
6942-10
I/O-Box 6753,
Inputs
brown (bn)
IN 1
– 8
+ 24 V
blue (bu)
IN 1 - 8
GND
black (bk)
Free inputs IN 1 - 8
Free inputs
IN 1-8
Note:
If a ground-switching proximity switch is to be connected to a device (eg I / O
box 6753) with positive-switching inputs, this can be done by interposing the
constant current module Type 6753-302 as a "pull up" b 24 V and
the corresponding input (see Fig. 1)
The digital inputs of the compact controller 6942-10 are configured as "posi-
tive switching" by default.
By changing the configuration (see Type 6744-10) each input can be
customized
also be switched to "mass switching".
Picture 1
Содержание MAK TIGER 3003
Страница 13: ...Safety precautions 5 Measuring System MAK TIGER 3003 Service manual SA 131128 10 03 2020...
Страница 31: ...Measuring system MAK TIGER 3003 23 Measuring System MAK TIGER 3003 Service manual SA 131128 10 03 2020...
Страница 32: ...Measuring system MAK TIGER 3003 24 Measuring System MAK TIGER 3003 Service manual SA 131128 10 03 2020...
Страница 187: ...ULTRASAMPLER sample collection system 179 Measuring System MAK TIGER 3003 Service manual SA 131128 10 03 2020 Dimensions...
Страница 216: ...Measuring System MAK TIGER 3003 Service manual SA 131128 10 03 2020...