7
Dimensions and Clearance Requirements
DIamond TII dimensions - 13.5" W X 18.5" D X 19.5" H (34.3 cm X 47.0 cm X 49.5 cm)
Clearances:
Sides - 9" (22.9 cm) minimum to allow for servicing
Above - 3" (7.6 cm) minimum for removal of top cover
Front - 4.75" (12.1 cm) minimum for opening front door
Electrical Requirements
The DIamond TII is equipped with two power cords and corresponding fuses taped to each power cord to be
plugged into a grounded electrical outlet of the appropriate voltage.
All models:
100-240 VAC +5% -10%, 47-63 Hz.
Feedwater Requirements
1
Types
1
Tap (Potable)
Turbidity
1.0 N.T.U. maximum
Pressure Range
2 bar (30 psig) to 6.9 bar (100 psig)
Temperature Range
4-40°C (39-104°F)
Minimum Inlet Flow Requirements
100 lph
pH
3-10
TDS
600 (max. ppm as CaCO3)
Silt Density Index
<5%
Free Chlorine
<3 ppm
Langlier Saturation Index
<1
Iron (total as Fe)
<0.5 ppm
Silica
<30 ppm
DI Product Water
Quality
Resistivity: ASTM Type II, NCCLS Type I >15 megohm/cm
TOC: <15 PPB
Flow Rate: up to 1 LPM with a new D3750 final filter
Bacteria: Less than 1 CFU/ml
Volumetric Dispense:
Accuracy: ±5%
Repeatability: ±3%
Feedwater suitability must be determined by a water analysis
General Specifications
1
500 ppm NaCl feedwater solution @ 25°C
(77°F), @ 4.5 bar (65 psig), 15% recovery.
Flow rate and recovery will decrease with
lower water temperature and pressure.
Barnstead recommends the use of a hot
water mixing valve before the RO system for
water cooler than 15°C (59°F).