14
Water Connection Details
Push-to-Connect Fitting Tubing
Installation (Refer to Fig. 2)
The following instructions will apply when you need to
attach a piece of tubing to your DIamond TII during instal-
lation, unless otherwise noted in the installation instruc-
tions. To make tubing connections:
1.
Make sure the tubing is cut off reasonably
square and that no plastic burrs or ridges
are present.
2.
Mark from end of tube an insertion length of 3/4”
(1.9 cm).
3.
Wet the tube end with water and insert the tube
straight into the fitting until it bottoms out on the
interior shoulder and the insertion mark is no
longer visible.
Push-to-Connect Fitting Tubing Removal
(Refer to Fig. 3)
1.
Using the tool provided (AYX23), push the collet
toward the body while pulling on the tubing to
release the tube.
I
NSTALLATION
COLLET
TUBING
TYPICAL
FITTING
3/4"
Insertion
Length
Figure 2
Tubing Installation
Note
For easier insertion, wet the end of
the tubing with water.
Note
Figure 2 and 3 will be used at the
connection to the atmospheric drain.
Figure 3
Tubing Removal
TUBE REMOVAL TOOL
TUBING
COLLET
TYPICAL FITTING
PLACE REMOVAL TOOL UP AGAINST
COLLET AND PUSH FORWARD WHILE
SIMULTANEOUSLY PULLING THE TUBE
BACK