20
8.
Make the PRODUCT (E) and REJECT 2 (F)
connections on left most membrane by first wet-
ting the tube stem of the elbow with water and
pushing it firmly into the connector. See Figure
9.
9. The door can remain open until the unit has
been operating and you have checked for possi-
ble water leaks at the membrane connections.
10. Rinse membranes per the instructions in the
“Initial Operation” section of this manual.
Cartridge Pack Installation
Cartridge packs will come bagged with four manifold
connection caps.
Each cartridge pack includes one 0.2 micron absolute
final filter.
1.
Open right side door. Remove the four caps
from the top of the cartridge pack.
2.
Verify that each of the four posts on the car-
tridge pack has an o-ring.
3.
Lift unit manifold upward, insert cartridge pack
and align the pack so that it mates with the unit
manifold.
4.
Lower unit manifold until it is flush with the top
of the cartridge pack.
5.
Hand tighten wing head screw securely.
6.
Close door.
7.
DO NOT
install the 0.2 micron filter and bell
assembly at this time.
I
NSTALLATION
Caution
Do not allow the DIamond TII to oper-
ate unless water is available to the
unit.
Figure 10: Dual Cartridge Pack
Figure 11: Cartridge Pack Installed
Note orientation of top cartridge
pack manifold