32
Note
When the RO operation is automati-
cally turned on, RO water will be put
into the reservoir. Anytime the RO
purity falls below the user-selected
% rejection setting the System Alarm
LED will illuminate.
Note
Volumetric dispensing is based upon
total fluid volume entering the
DIamond TII. Point of dispensing accu-
racy will be compromised if you are
dispensing water volumetrically while
simultaneously dispensing water man-
ually through the remote dispenser
accessory.
Automatic Dispensing
This feature is accessible from any screen when the volu-
metric dispensing icon is an available switch.
1.
Place a container under the dispenser.
2.
Press the volumetric dispensing icon.
The screen will show current settings
for Remote Volumetric, Unit Volumetric, Timed
Dispense or OFF. Press MODE to choose set-
ting and use the up and down arrows to set vol-
ume or time.
Timed Dispensing
UP arrow: increments by 5 minutes
DOWN arrow: Decrement by 1 minute
Maximum: 30 or 60 minutes depending on reservoir
volume
Minimum: 1 minute
Volumetric Dispensing
UP arrow: 250ml up to 1 liter, then increments of 5L
DOWN arrow: Decrement by 250 ml up to 1L, then decre-
ment by 1L
Maximum: 30L or 60L depending on reservoir volume
Minimum: 250 ml
Reservoir Replenishment
If the water level is below 75%, the external reservoir will
begin automatically refilling in either RUN mode or
STANDBY mode by the self-contained RO system. The
water level in the reservoir is determined by internal sen-
sors in the reservoir. Once the water level fills up to the
full level, the RO system will enter STANDBY mode.
I
NITIAL
O
PERATION
Note
Once Volumetric or Timed Dispensing
values are set, they will remain in
memory until changed by the user. To
turn the dispense method OFF, select
“OFF” in the “Dispense Method”
menu.