ROUBLESHOOTING
Routine Care
Message
Possible Causes
Action
Replace Carbon
Chlorine capacity of the carbon
Replace Carbon Prefilter
prefilter has been reached, or 1 year
(D502115) and reset the carbon
service life has expired.
prefilter timer.
Check Prefilter
System has sensed that the prefilter
Replace the RO Prefilter
is plugged.
(D502113) and reset the prefilter
timer.
Replace Prefilter
1 year service life of the prefilter has
Replace the RO Prefilter
expired.
(D502113) and reset the prefilter
timer.
Replace Membrane
1 year service life of the membrane(s)
Replace the RO membrane(s)
has expired.
(FLX1265X1) and reset the
membrane timer.
Replace Res Filter
6 month service life of vent filter
Replace vent filter element (25001)
has expired.
and reset the vent filter timer.
Replace MPS Cart
1 year service life of the MPS
Replace the MPS cartridge
cartridge has expired.
(D502114) and reset the MPS
cartridge timer.
Replace DI Pack
1 year service life of the cartridge
Sanitize the system, replace the
pack has expired.
cartridge pack (D502137), and
reset the sanitization and cartridge
pack timers.
Replace System UV
6 month peak service life of the
Replace the system UV lamp
system UV lamp has expired.
(LMX13) and reset the system UV
lamp timer.
Replace Res UV
1 year peak service life of the
Replace the reservoir UV lamp
system UV lamp has expired.
(LMX31) and reset the reservoir
UV lamp timer.
Sanitize System
1 year system sanitization reminder
Sanitize the system and reset
has expired.
the sanitization timer.