13
Component Installation
UV Lamp Installation (See Fig. 7)
Locate the UV chamber inside the left door of the
DIamond TII. Install the UV Lamp as follows:
1.
Disconnect the power cord from the unit.
2.
Access the left side of the unit by unlatching the
screw securing the door with a flathead screw-
driver. Open the door.
3.
Remove the UV lamp from its packaging. DO
NOT TOUCH THE GLASS PORTION OF THE
LAMP! It is recommended that lint-free gloves
be worn when handling the lamp. The glass por-
tion must be free of fingerprints, perspiration,
etc. Even a light coating of perspiration will
reduce the effectiveness of the lamp. If the glass
portion of the lamp is touched, clean it with a
damp, lint-free cloth: use isopropyl alcohol as
required.
4.
Remove black cap by sliding it off of UV cham-
ber.
5.
Insert the lamp halfway into the UV lamp cham-
ber and plug it in, then fully insert the lamp into
the chamber, and replace black cap.
6.
Make sure the UV lamp cable is tucked behind
the chamber before closing and relatching the
door.
7.
Reset the UV lamp timer as directed in the
“Resetting Routine Timers” section of this manu-
al.
I
NSTALLATION
Caution
DO NOT TOUCH THE GLASS POR-
TION OF THE LAMP! It is recom-
mended that lint-free gloves be worn
when handling the lamp. The glass
portion must be free of fingerprints,
perspiration, etc. Even a light coating
of perspiration will reduce the effec-
tiveness of the lamp. If the glass por-
tion of the lamp is touched, clean it
with a damp, lint-free cloth: use
isopropyl alcohol as required.
Note
The UV lamp contains mercury. If
broken or no longer needed, do not
dispose of the UV lamp in the trash.
Recycle or dispose of the UV lamp as
hazardous waste.