33
User Settings
Setting the Displayed Language
1.
Upon power up, press any of the four buttons
below the display when the display shows “LAN-
GUAGE.”
Or, if the unit’s power is already on:
1.
Press MENU and scroll to the LANGUAGE
screen and press OK.
2.
Press the UP or DOWN arrow to select the lan-
guage desired and press OK.
Setting the Time and Date
1.
Press MENU and scroll to the SET TIME screen
and press OK. Using the arrow keys, select time
and press OK.
2.
Scroll to the SET DATE screen and press OK.
Using the arrow keys, select date and press OK.
Setting the Low Product Purity Alarm
The Low Product Purity Alarm monitors the purity of the
product water at the system dispense point. The alarm is
adjustable from 1.0 m
Ω
-cm (1.0µS/cm) to 15.0 M
Ω
-cm
(.067µS/cm), the factory default is 1.0 M
Ω
-cm. To adjust
the alarm:
1.
Press MENU. Scroll to DI Options and choose
Purity Alarm.
2.
Use the up and down arrows to set the low
alarm. Press OK to accept settings.
The alarm will alert the user when the product purity falls
below the product purity alarm level by illuminating the
SYSTEM ALARM LED. See details by pressing MENU
and choosing System Alarm. See the “Error Conditions”
section later in this manual to interpret the error.
I
NITIAL
O
PERATION
Note
Time format is 12 hour, and date for-
mat is day/month/year.