Cleaning and care
Thanks to its smooth surfaces and rounded corners the Barkey S-line is simple and easy to clean.
Clean the surfaces of the control unit and the fastening clamp, as well as the inside and outside of
the warming section, and wipe them with disinfectant regularly, where possible after each use.
WARNING
Before cleaning or wiping with disinfectant:
•
Switch the Barkey S-line off by pressing the on/off button and pull the device's
mains plug out of the mains socket.
•
Do not apply cleaners and disinfectants directly to the device. Moisten a cloth
with them and use this cloth on the device.
•
Never put cleaning liquids into the device's ventilation openings.
•
Do not disinfect the device with steam (eg in an autoclave), hot air or with
thermochemical agents.
Commercially available preparations on an alcohol base without oxygen-releasing components
(with sodium hypochlorite with a 1.4% bleach solution or aldehyde content <0.2%) should be used
as disinfectants.
►
Clean the device where possible with a water/soap solution.
►
Dry the surfaces and use a surface disinfectant.
►
Observe the reaction time indicated by the disinfectant manufacturer!
►
After expiry of the reaction time dry the device with disposable cloths.
WARNING
The operator may not use any cleaning or decontamination methods other than those recommended
by the manufacturer.
If other cleaning or decontamination methods are to be used, a check must be made beforehand
with the manufacturer to see whether these suggested methods damage the device.