Safety information
10
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe - Germany - Telephone: +49(0)5202-9801-0
Service department: +49(0)5202-9801-30 Fax: +49(0)5202-9801-99 - email: [email protected]
Instructions for Use
Barkey S-line
3
Safety information
3.1
Safety information for using the device
WARNING
Check the device for damage before each use. Damaged de-
vices may not be used.
In order not to endanger the patient or the operator, do not use
the device if one of the following conditions is found:
-
damaged or worn mains cable, mains plug, control lead
-
damaged housing, damaged control panel
-
device was dropped, or the internal electronic components
have become wet
-
device does not warm correctly (shown by optical and
acoustic alarms)
-
yellow alarm lamp or acoustic alarm is activated
All electrical installations must conform to the respective valid
regulations and standards and the specifications stipulated by
the manufacturer. The owner is responsible for the correct in-
stallation.
Read these instructions for use carefully before using the de-
vice.
The manufacturer does not assume any liability for the patient's
safety if the operating, maintenance and calibration procedures
for the system do not conform to the procedures in the instruc-
tions and in the service manual.
Take account of the requirements for the qualifications of the
device operators. The operating personnel must be suitably
trained and qualified.
It is imperative that you read the information on the use for the
intended purpose and the safety information in these instruc-
tions for use.
Use the device solely for the intended purpose and in confor-
mity with the procedures described in these instructions for
use.
Before being used the Barkey S-line must be acclimatised to
the ambient temperature at the location of use for at least one
hour.
If infusion solutions or medicines are warmed or kept warm,
the operator must ensure that they are suitable for the preset
temperature in the device and retain their efficacy.
Turn the device's fixing screw so tight that the holder is fixed
securely to the infusion stand. It must not be possible to shake
or slide it! If the holder is not fixed securely, the device may
slide down with the infusion line and prevent or interrupt the
infusion flow.
The warming device can be used at ambient temperatures of
+10°C to +40°C and is approved for permanent use.