Installation / start-up
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe - Germany - Telephone: +49(0)5202-9801-0
19
Service department: +49(0)5202-9801-30 Fax: +49 (0)5202-9801-99 - email: [email protected]
Instructions for Use
Barkey S-line
4
Installation / start-up
4.1
Location
The Barkey S-line is conceived for use in buildings as a fixed installation. Mobile use is not
planned.
Before operating the device, check the location for suitability.
WARNING
When selecting the installation location, it is essential to take account of the proximity to other
devices with electromagnetic emissions indicated in these instructions for use. The Barkey S-line is
not approved for use in potentially explosive areas.
It is also essential that you take account of the information in these instructions for use that indi-
cates the environmental conditions in which the device may not be used or in which its function
can be disturbed.
The device is ventilated from the back. The ventilation slots on the back of the device must not be
covered.
Fasten the Barkey S-line to an infusion stand as shown in chapter 4.2.1. The control panel must be
easily accessible, the device operator must be able to read the display and the lamps.
To achieve the maximum benefit, position the Barkey S-line so that the end of the warming section
is close to the injection point on the patient's body.
4.2
Starting installation / start-up
The Barkey S-line is designed for the mains voltage range shown on the rating plate and must be
fastened to an infusion stand.
WARNING
Before the installation and start-up of the device, read the instructions for use carefully in full.
All electrical installations must conform to the currently valid regulations and standards and the
specifications stipulated by the manufacturer. The owner is responsible for the correct installation..
Always check the device for damage before each use. Do not use a damaged device. In order to
safeguard the safety of the patient and/or the operator, do not use the device if one of the following
conditions is found:
•
damaged or worn mains cable, mains plug, control lead
•
damaged housing, damaged control panel
•
device was dropped, or the internal electronic components have become
wet