Device description
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe - Germany - Telephone: +49(0)5202-9801-0
9
Service department: +49(0)5202-9801-30 Fax: +49 (0)5202-9801-99 - email: [email protected]
Instructions for Use
Barkey S-line
2.9
Cooling the control unit
The control unit is fitted with ventilation slots on the back.
ATTENTION
The ventilation slots must not be covered or closed.
There is a fan behind the lower ventilation slots that cools the electronics. The fan is controlled in
dependence on the temperature, i.e. it can be switched off fully at low temperatures or run at full
speed at high temperatures of the electronics.
The fan's noise level is very low, so that it does not create noticeable disturbance.
2.10
Overtemperature protector
An independent overtemperature protector monitors the temperature in the warming section.
In case of a fault, if a temperature 43°C +/- 1°C is reached, the warming section's heater is switched
off, the luminous ring in the display and control panel shows yellow and an acoustic signal is
heard.
In this case switch the device off or pull the mains plug out and wait until the warming section has
cooled down. The cooling down process can take some minutes. After this you can switch device
on again. However, if the cause of the fault has not been eliminated, the fault will occur again and
the device must be inspected by a trained service technician.
WARNING
On a fault signal the infusion fluid to the patient must be stopped immediately or the warming de-
vice removed from the patient.
The Barkey S-line may not be used in case of a device fault. Barkey GmbH & Co. KG or author-
ised personnel must inspect the device.
2.11
Fault signal
On a fault signal the luminous ring shows yellow and an acoustic signal is heard. The fault signal
occurs if
-
the overtemperature protector has triggered (see chapter 2.9), or
-
one of the warming conductors integrated in the section is defective, or
-
a sensor or a sensor cable is broken, or
-
the temperature of the electronics was found to be too high.
If a fault signal occurs, switch the device off or pull the mains plug out. Wait about 5 minutes until
the warming section has cooled down. Switch the device on again. If the fault occurs again, the
device must be inspected by a trained service technician.
WARNING
On a fault signal the infusion of fluid to the patient must be stopped immediately or the warming
device removed from the patient.
The Barkey S-line may not be used in case of a device fault. Barkey GmbH & Co. KG or author-
ised personnel must inspect the device.