Bante900P Portable Multiparameter Water Quality Meter 

Instruction Manual 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

BANTE INSTRUMENTS LIMITED 

Technology to create the future...

.

 

 

Содержание 900P

Страница 1: ...Bante900P Portable Multiparameter Water Quality Meter Instruction Manual BANTE INSTRUMENTS LIMITED Technology to create the future...

Страница 2: ...No potential electromagnetic interference The following list describes the standard accessories of the meter After the unpacking please check all accessories are complete If any are damaged or missing...

Страница 3: ...battery is depleted the indicator will disappear Calibration mode indicator Indicates meter is in the calibration mode Stable indicator Indicates the measured value has stabilized Setup mode indicato...

Страница 4: ...ivity TDS Salinity Resistivity DO Press and hold the key to enter temperature setting mode CAL Press the key to enter the calibration mode Press and hold the key to enter the setup menu MI Press the k...

Страница 5: ...robe 3 6 pin Connector For connecting the conductivity electrode or dissolved oxygen probe Inserting the Batteries Before using the meter insert three 1 5V AA alkaline batteries into battery compartme...

Страница 6: ...BNC connector into corresponding connector socket Rotate and push the connector clockwise until it locks FOR THE CONDUCTIVITY ELECTRODE OR DISSOLVED OXYGEN PROBE Insert the 6 pin connector into corres...

Страница 7: ...issolved Preparation of pH4 01 and 10 01 standard buffer solutions are the same as above Prepared standard buffer solutions should be stored in hermetically sealed glass containers Refilling Electroly...

Страница 8: ...e diluted standard solution e g 100ppm for at least 15 minutes FOR THE CONDUCTIVITY ELECTRODE Soak the electrode for a few minutes in tap water to remove dirt and oil stains on the electrode FOR THE D...

Страница 9: ...10 01 12 45 NIST 1 68 4 01 6 96 9 18 12 45 DIN 1 09 4 65 6 79 9 23 12 75 User Defined Buffer Calibration Points Select the number of calibration points you intend to calibrate the meter 1 point 2 poi...

Страница 10: ...ient Select the temperature coefficient of sample Setting Range 0 0 to 10 0 o C 2 10 Calibration Points Select the number of calibration points 1 point 2 points 3 points 4 points 5 points Pure Water C...

Страница 11: ...grees Fahrenheit mg L Milligrams per liter ppm Parts per million mmHg Pressure unit kPa Pressure unit GENERAL OPTIONS PARAMETER DESCRIPTION OPTIONS DESCRIPTION DEFAULT Stable Criteria Sets the stabili...

Страница 12: ...Clear all stored data Enable Disable Reset Reset function allows user to restore the meter back to factory default parameters When this function is enabled all calibration values and selected paramete...

Страница 13: ...s ENTER key to confirm the display shows an option in the submenu 5 Press or key to select the desired option 6 Press ENTER key to confirm the meter returns to measurement mode Setting is completed EX...

Страница 14: ...a Follow the steps below to set the date and time during the first use 1 Press and hold the key for 3 seconds to enter the setup menu 2 Press or key until the display shows DATE option 3 Press ENTER k...

Страница 15: ...the display the meter is now switched to automatic temperature compensation mode MANUAL TEMPERATURE COMPENSATION 1 DO NOT connect the temperature probe to meter 2 Press and hold the o C key for 3 seco...

Страница 16: ...9 otherwise calibration will not be accepted Make sure to calibrate the meter when attaching a new electrode or during first use DO NOT reuse calibration solution after calibration contaminants in sol...

Страница 17: ...int calibration 2 3 Rinse the pH electrode with distilled water Dip the electrode into the next standard buffer solution e g pH4 01 The meter automatically senses the current calibration solution and...

Страница 18: ...ay shows CAL2 The meter prompts you to continue with second point calibration 3 5 Rinse the pH electrode with distilled water Dip the electrode into the next custom buffer solution 3 6 Press or key to...

Страница 19: ...pH electrode for diagnostic purposes 1 Press MR key in the pH measurement mode the display shows LOC P 01 2 Press or key until the meter shows ELE P 02 Electrode Diagnosis 3 Press ENTER key to confirm...

Страница 20: ...ss key again the meter shows the slope of each pH buffer group 6 After the browsing press MEAS key to exit the currect mode If you are not calibrate the meter or calibration is not successfully press...

Страница 21: ...e is needed 1 Press MODE key until the meter shows indicator 2 Rinse the ORP electrode with distilled water Dip the electrode into the calibration solution Stir the sensor gently 3 Press CAL key the m...

Страница 22: ...ss MR key in the ORP measurement mode the display shows LOC P 01 2 Press or key until the meter shows ELE P 02 Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration dat...

Страница 23: ...ate measurement results we recommend that you perform ion calibration and measurement at same temperature If you are not calibrate the meter or calibration is not successfully the display will always...

Страница 24: ...lective Electrode with distilled water then rinse with a small amount of ion standard solution 6 Dip the electrode into corresponding calibration solution Stir the sensor gently to create a homogenous...

Страница 25: ...egins flashing Wait for the measured value to stabilize the display shows 1 CAL3 The meter prompts you to continue with third point calibration 10 Repeat steps 8 to 9 above until the display shows END...

Страница 26: ...ey in the ion measurement mode the meter shows LOC P 01 2 Press or key until the display shows ELE P 02 Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date Forma...

Страница 27: ...to 5 points calibration To ensure higher accuracy we recommend that you perform 3 points calibration or select a standard value close to the sample value you are measuring The meter will automaticall...

Страница 28: ...ation solution the meter automatically shows current calibration standard e g 1413 S cm 1 5 Press ENTER key to confirm Calibration indicator begins flashing 1 6 Wait for the measured value to stabiliz...

Страница 29: ...ion solution the meter automatically shows current calibration standard e g 12 88mS cm 2 4 Press ENTER key to confirm Calibration indicator begins flashing 2 5 Wait for the measured value to stabilize...

Страница 30: ...er Quality Meter 29 3 2 Press and hold the key for 3 seconds to enter setup menu the display shows CELL option 3 3 Press ENTER key to confirm the meter enters the cell constant setting mode 3 4 Press...

Страница 31: ...s ENTER key to confirm until the setting values stop flashing The meter returns to measurement mode automatically Setting is completed During the calibration process if you want to exit calibration pr...

Страница 32: ...onductivity measurement mode the display shows LOC P 01 2 Press or key until the meter shows ELE P 02 Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date Format...

Страница 33: ...ation in the air For the 2 points calibration you need to use saturated anhydrous sodium sulfite solution zero oxygen solution 100 SATURATION CALIBRATION 1 1 Press MODE key until the meter shows indic...

Страница 34: ...1 2 3 Press or key until the display shows 0 CAL1 2 4 Dip the dissolved oxygen probe into the saturated anhydrous sodium sulfite solution zero oxygen solution Stir the probe gently 2 5 Press ENTER key...

Страница 35: ...alibration will simultaneously calibrate the corresponding mg L or ppm concentration value Therefore additional mg L calibration isn t required in most circumstances DO Calibration in mg L or ppm Mode...

Страница 36: ...begins flashing 1 7 Wait for the measured value to stabilize the display will show END The meter returns to measurement mode automatically Single point calibration is completed 2 POINTS CALIBRATION 2...

Страница 37: ...the display shows 8 25 CAL2 The meter prompts you to continue with second point calibration 2 7 Dip the dissolved oxygen probe into air saturated water for 3 to 5 minutes press ENTER key to confirm Wa...

Страница 38: ...enabled if the temperature reading displayed differs from that of an accurate thermometer you need to calibrate the meter 1 Press and hold the o C key for 3 seconds to enter temperature calibration mo...

Страница 39: ...alue on the display ORP Measurement Bante900P portable meter contains two millivolt measurement modes ABSOLUTE MILLIVOLT MODE Press MODE key until the display shows measurement unit mV the meter is no...

Страница 40: ...TDS or Resistivity indicator 2 Rinse the conductivity electrode thoroughly with distilled water 3 Dip the electrode into the sample solution stir the sensor gently 4 Wait for the reading to stabilize...

Страница 41: ...d damaged or eroded 1 Press MODE key until the meter shows indicator 2 Connecting the dissolved oxygen probe to meter wait for 15 minutes to polarize the sensor 3 If necessary to set the atmospheric p...

Страница 42: ...500 data sets to be stored and recalled MEMORY INPUT During the measurement process press MI key to input measured value into the memory Memory indicator appears on the display MEMORY RECALL 1 Press M...

Страница 43: ...meter Water Quality Meter 42 3 Press key the meter shows date and time of the stored data Format mm dd hh mm 4 Press key again the display shows the stored data 5 After the browsing press MEAS key to...

Страница 44: ...on on the start bar the DAS splash screen appears for few seconds and then the application starts RECEIVING THE STORED DATA Connect the USB cable to meter and computer communication port COM1 baud rat...

Страница 45: ...on the data field then click the Delete button Note Once deleted the data cannot be recovered GRAPH MODE The graph mode will help user to view variations of the measured parameter continuously Click t...

Страница 46: ...cel file C Program Files Data Acquisition System Excel NOTE The software is run in operating system Windows XP If your system is Windows 7 please using the file CP210x_VCP_Windows to update your devic...

Страница 47: ...tes If this fails the electrode requires activation Soak the pH electrode in 0 1M HCl for 5 minutes Remove and rinse with deionized water then place in 0 1M NaOH for 5 minutes Remove and rinse again a...

Страница 48: ...e should be removed very carefully with a cotton bud soaked in solvent taking care not to touch the metal parts of the inner cell DO Probe Care and Maintenance Please always keep probe s membrane mois...

Страница 49: ...e solution Zero oxygen solution is contaminated Replace the calibration solution pH electrode is out of service life Replace the pH electrode Keypad is not working properly Replace the batteries Speci...

Страница 50: ...ax 200ppt depending on factor setting Accuracy 1 F S TDS Factor 0 1 1 0 Default 0 5 Salinity Range 0 10ppt Max 80ppt Accuracy 1 F S Measurement Modes Seawater or Practical Salinity Resistivity Range 0...

Страница 51: ...bility Conditions Low or High Calibration Due 0 to 31 days Power Off Manual or Automatic 10 20 30 minutes Reset Function Yes Memory Stores up to 500 data sets Output USB Communication Interface Connec...

Страница 52: ...t can not fully meet your measurement requirements please refer to the table below to select an applicable sensor SAMPLE TYPE P11 P12 P13 P14 P16 P17 P18 P19 P20 P21 Agar Beer Blood Products Bread Dou...

Страница 53: ...Dissolve 22 365 grams of potassium chloride reagent KCL in the 100mL distilled water 2 Pour the 50mL of KCL solution into the beaker 3 Pour the 50mL of pH4 01 standard buffer solution into the same b...

Страница 54: ...o same sample solution B Make sure both sample solutions have about 5o C to 10o C difference 6 Record the temperature value TB and conductivity value CTB 7 Calculate the temperature coefficient of sam...

Страница 55: ...800 91 7 688 900 90 5 679 1000 89 4 671 1100 88 3 662 1200 87 2 654 1300 86 1 646 1400 85 0 638 1500 84 0 630 1600 82 9 622 1700 81 9 614 1800 80 9 607 1900 79 9 599 2000 78 9 592 Addendum 8 How to pr...

Страница 56: ...hese efforts and we expect to have little or no such materials in our product in the coming years We welcome our customer suggestions on how to speed up these efforts Warranty The warranty period for...