2 Preparation
Carefully unpack the ultrasound generator, ultrasonic converter and accessories, and inspect
them for completeness or possible transportation damage. If damage or defects are ascertained,
notify this to the forwarding agent and the supplier immediately and in writing.
Before start-up, the ultrasonic homogeniser should be allowed to stand for two hours at its
operating location so that it can adjust to the ambient conditions.
2.1
Setup / assembly
Place
the ultrasound generator atop a
fi
rm, level and dry surface. In doing so,
•
do not group it or stack it over other electrical or electronic devices,
•
do not cover the ventilation holes on the ultrasonic converter’s housing,
•
guard against moisture and wetness – risk of electric shock.
The
ultrasound generator must be positioned in such a way that disconnection of the power
supply is easy.
The
ultrasonic converter is connected to the ultrasound generator via a push-pull circular
connector with lock. To connect, hold the mains connector from the handle and position it
in such a way that the arrow points toward the left. The mains connector can only be plugged
in when in this position. Plug the mains connector into the socket up to the limit, and let it snap
into place. In order to unplug the connection, hold the mains connector only from the handle,
pull it back and remove the connector. If necessary, see chapter 1.
Verify that the mains switch is in the “0” position.
Before the mains connection, identify the mains voltage at hand and compare
it with the voltage speci
fi
ed on the type plate on the generator bottom.
Connect to a grounded socket only if type is suitable.
2.2
Putting into service
Connect the ultrasound generator to the mains ( earthed socket) and switch on.
Adjust the sieve.