10 / 33
51826-001 en/2023-01
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
1.4.5
Remote operation
Several options are available for remote control and remote operation.
Depending on requirements, the most convenient solution may be selected.
Operating
element
Method
Functions
Connection
1
Key
manual
Ultrasonic operation ON/OFF
Fixed on the ultrasonic
converter
2
Foot switch
foot-operated Ultrasonic operation ON/OFF
Remote control
connector
3
External
control signal • Ultrasonic operation ON/OFF
• Status con
fi
rmation
Remote control
connector
4
External
data protocol • Status inquiry
• Full device control
RS 232 interface
a) Remote control connector
Contact assignment and functional description of the remote control connector
Diagram
Contact
Assignment Signal
Function
1
2
3
4
5
1
Input
0V L
5V H
L
Ultrasound operation ON
H
Ultrasound operation OFF
2
Output
0V L
5V H
L
Ultrasound operation OFF
H
Ultrasound operation ON
3
Earth
4
Output
like contact 2
5
n.a.
Notes:
• The connection of power or voltage sources of any type is not permitted.
• The source resistance for the outputs is of 3.3 k
Ω
. The input resistance for subsequent
signal inputs, e.g. optocouplers, must be greater than or equal to 47 k
Ω
.
• The function of the signal input (contact 1) depends on the operating mode that is set for
the pulsation (see chapter 3.2.2). With the operating mode “by hand key”, control is performed
statically, i.e. the ultrasonic operation is always activated under L status, and deactivated
under H status.
With all other operating modes, the input works as an edge-controlled changeover switch.
The input resistance totals approx. 50 k
Ω
.
• When connecting a changeover contact (toggle function) to connector 1, the make time
must be > 100 ms.