By pressing the “SET” key, the content of the working memory (current program) can be displayed
for comparison purposes, e.g. to facilitate the selection of a memory location.
Use the
and
arrow keys to select the desired memory location for loading or saving
the sonication program.
Loading the sonication programme
In order to copy the content of the selected data memory in the working memory, the arrow
key is pressed and held until an audible signal is emitted. The data can be used after leaving
the program administration.
Saving the sonication programme
By continuously pressing the arrow key until an audible signal is emitted, the content of
the working memory is transferred into the selected data memory. Then, the new contents
are displayed.
SET
select Program
No 1
Amplitude
10%
Time set
00:01:00
Pulse
4.0s
4.0s
Temp alarm at
0°
load
[
]* store [
]*
quit
[start]
No 5
The “START/STOP” key is used to exit memory management and return to stand-by mode.
3.4
Activating / deactivating the ultrasound
Ultrasonic operation can be started and stopped by pressing the “START/STOP” key or the key on
the ultrasonic converter. With an activated target time, the ultrasound operation ends and a signal
tone (short + short) is emitted once the target time has elapsed. When the device is restarted,
the displays for the elapsed sonication time and for the energy are set to zero.
Prog 5* Ampl
20%
Time set
00:01:00
00:00:09
Pulsation
off
Energy 0.000
kJ
Elapsed
Sonication time
Actual value
Set value
Energy
In contrast to stand-by mode, ultrasound operation displays the actually achieved value in bar
graph form, in addition to the pre-set amplitude or power.
Note:
Depending upon the physical conditions, i.e. the viscosity of the sonicated medium and
the resulting control processes, a di
ff
erence between the target and the actual value, as well as
a
fl
uctuating display, may occur.