BANDELIN SONOPULS HD 4200-SB Скачать руководство пользователя страница 29

29 / 33

51826-001 en/2023-01

1

 electronic GmbH & Co. KG  •  Heinrichstraße 3-4  •  12207 Berlin  •  Deutschland  •  [email protected]

6.3  

Troubleshooting  

Errors may arise 

•  on the plug connections 

•  on the  ultrasonic converter

•  on the  ultrasound generator 

The device is robustly constructed and designed for a high level of reliability. Nevertheless, 
the possibility of a malfunction due to a  defective component can never be fully discounted. 
Mechanical  defects of the connection for the ultrasonic converter, the plug connectors, 
the  ultrasonic converter, etc., are possible as a result of frequent  use or even incorrect handling, 
e.g. by dropping them.

Once detected by the device, signi

fi

 cant errors are signalled using a red LED and an intermittent 

signal tone (3 times), and displayed with an error number – see the following table. 

Error No.

Possible cause

Remedial action

Converter: 
No USC

No ultrasonic converter (UW) connected 
to the ultrasound generator (GM)

Connect the UW to the horn/probe on the GM. 
The GM should then automatically recognise 
which UW has been plugged in.

002

Frequency setting not possible

Perform frequency scan. If the error occurs again, 
contact BANDELIN.

003

No power output, amplitude setting 
not possible

Error correction should only be carried out by 
BANDELIN. Send in the complete ultrasonic 
homogeniser.

004

Insu

  cient mains voltage

Check mains connection, check fuse.

010

Frequency synchronisation 
malfunctioning, frequency display 
is currently not possible

-   Check whether the probe has been mounted 

correctly = su

  ciently tight

-  Check for damage:

•  Horn/probe contact surfaces
• Horn/probe 

threads

•  Probe radiation surface

If needed, replace the UW/probe and then 
check its function. Use the “probe check” or 
“scan frequency” functions for this purpose. 
After successfully completing the frequency scan, 
do not restart operation until initialisation is complete. 
In the event of failure, contact BANDELIN.

011

No return signal from UW, UW 
or probe defective

012

Error in frequency scan, no resonance 
frequency found, probe defective

014

Permissible internal operating 
temperature exceeded, ultrasonic 
homogeniser switches o

 

Allow the ultrasonic homogeniser to cool down 
for at least 10 minutes; if necessary, check 
with a temperature sensor that the operating 
temperature of the GM has dropped to < 50°C.

020

Unknown command (communication 
error), remote control command was not 
executed or ignored

Observe the application information “Protocol and 
command set for remote operation” and check 
commands.

021

Incorrect command length 
(communication error), remote control 
command was not executed or ignored

022

Unknown type (e.g. probe)

Check the sieve, UW, standard/booster horn, 
and probe for compatibility.

Содержание SONOPULS HD 4200-SB

Страница 1: ...7 Ultrasonic Homogeniser HD 4200 SB 51826 001 en 2023 01 Operating Instructions...

Страница 2: ...original version will take precedence BANDELIN accepts no responsibility or liability for damage caused by improper handling or usage contrary to the intended purpose The documentation has been prepa...

Страница 3: ...event of improper handling or usage contrary to the intended purpose In the event of unauthorised alterations modifications both the warranty claim and the CE conformity will become void If service is...

Страница 4: ...Electromagnetic compatibility EMC 9 1 4 5 Remote operation 10 1 5 Warnings and safety precautions 12 2 Preparation 14 2 1 Setup assembly 14 2 2 Putting into service 14 3 Operation 15 3 1 Operating ele...

Страница 5: ...homogeniser 26 5 1 Cleaning and care 26 5 2 Warehousing storage 26 6 Maintenance and repair 27 6 1 Maintenance 27 6 2 Functional checks 27 6 2 1 Testing the ultrasonic converter probe check Error 011...

Страница 6: ...onents can be modified with various accessories and adapted to specific applications The type specification and serial number are found on the type plate Product features Ultrasonic generator 1 in dur...

Страница 7: ...AMPLICHRON circuitry independently from the applied load as long as the maximum power allowed is not exceeded Thanks to these measures the reproducibility of the process parameters is guaranteed and v...

Страница 8: ...30 pursuant to DIN EN 60529 Generator Generator GM 4200 SB Power 2 maximum 70 W Ultrasonic frequency 20 kHz Weight approx 3 2 kg External dimensions L W H 335 150 230 mm 1 Frequency control The ultras...

Страница 9: ...IP 20 pursuant to DIN EN 60529 1 4 3 Additional data Ambient conditions Overvoltage category II Degree of contamination 2 Permissible relative humidity up to 31 C 80 Permissible relative humidity up...

Страница 10: ...Contact Assignment Signal Function 1 2 3 4 5 1 Input 0V L 5V H L Ultrasound operation ON H Ultrasound operation OFF 2 Output 0V L 5V H L Ultrasound operation OFF H Ultrasound operation ON 3 Earth 4 O...

Страница 11: ...esponsibility of the operator and is not supported by the manufacturer The manufacturer only guarantees the proper functioning of the interface The instruction set for the RS 232 interface will be mad...

Страница 12: ...arately Ultrasonic homogenisers adhere to prescribed EMC limit values so that it can be assumed that the electromagnetic radiation emanating from the devices is harmless to humans A binding statement...

Страница 13: ...conditions may result in EMC disruptions that may a ect devices that are in the direct vicinity During operation portable or mobile HF communication systems in the vicinity of the ultrasonic homogenis...

Страница 14: ...s risk of electric shock The ultrasound generator must be positioned in such a way that disconnection of the power supply is easy The ultrasonic converter is connected to the ultrasound generator via...

Страница 15: ...topping the ultrasound emission ending functions exiting the menu 3 SET key Call up of the menu Navigation within the menu 4 Arrow keys Setting the menu or operating parameters 5 Control LED 1 5 3 4 2...

Страница 16: ...ed If the ultrasonic converter was exchanged in the meantime the earlier settings will be replaced by standard values and must be re edited The temperature display is only visible when a temperature s...

Страница 17: ...eter has number values that can be edited the editing mode is then activated with the arrow key In the marked area inverse view of the parameter the desired value can then be set with the arrow keys I...

Страница 18: ...Stop at 0 C C Energy 0 000 kJ Time set nonstop b Setting the sonication time If a value of 59 minutes or seconds is exceeded when setting the time or if a shortfall occurs when setting back the time...

Страница 19: ...ontinuous sound manual manual pulsation can be performed with the button on the ultrasonic converter Ultrasound operation is active as long as the button on the ultrasonic converter is pressed a Setti...

Страница 20: ...ion is restored if no device fault or other cause is present The default value is o In order to activate the function set the parameter to start using the arrow keys and then press the START STOP key...

Страница 21: ...uctuations are balanced only on average Switch display display Here you can choose whether the cumulative energy or the current ultrasonic frequency is displayed 3 3 Loading saving sonication programm...

Страница 22: ...1 Amplitude 10 Time set 00 01 00 Pulse 4 0s 4 0s Temp alarm at 0 load store quit start No 5 The START STOP key is used to exit memory management and return to stand by mode 3 4 Activating deactivatin...

Страница 23: ...start again at 0 00 00 If the maximum displayable energy value of 99999 99 kJ is exceeded the display starts again at 0 000 kJ The display value for the energy is a guideline value and does not corres...

Страница 24: ...ean the sieve and inspect it at regular intervals for possible damage 4 2 1 Mounting the ultrasonic converter on the sieve Screw the ultrasonic converter and the sieve together firmly at the threaded...

Страница 25: ...pling between the ultrasonic converter and the sieve the stability of the cover and the loading of the sieve also play a role As such it is advisable to proceed as follows in order to prevent potentia...

Страница 26: ...The cleaning and disinfection if needed must be carried out regularly by the user with a suitable cleaning agent and a lint free cloth In most cases a cloth soaked in alcohol is su cient to wipe o the...

Страница 27: ...ation 6 2 Functional checks See also chapter 3 2 3 for description and operation For error messages Error 011 and Error 012 always start with the test according to chapter 6 2 1 first 6 2 1 Testing th...

Страница 28: ...T SET START STOP probe check off scan frequency off contrast 22 operating mode Power batch mode off 01 02 2014 08 20 on probe check off scan frequency off contrast 22 operating mode Power batch mode o...

Страница 29: ...ried out by BANDELIN Send in the complete ultrasonic homogeniser 004 Insu cient mains voltage Check mains connection check fuse 010 Frequency synchronisation malfunctioning frequency display is curren...

Страница 30: ...and sieve not secure Clean the surfaces and screw them back together tightly Sieve defective Check sieve for cracks disassemble and replace if necessary Make sure surfaces are clean and flat Liquid be...

Страница 31: ...ts The black housing of the ultrasonic converter must not be twisted against the coupling piece to the sieve The ultrasonic oscillating system and its electrical connections would be damaged as a resu...

Страница 32: ...able for this ultrasound application 9 Taking the unit out of service The device must be disposed of appropriately not with household waste Disposal must be conducted in accordance with the Waste Elec...

Страница 33: ...BANDELIN electronic GmbH Co KG Heinrichstra e 3 4 12207 Berlin Deutschland www bandelin com info bandelin com 49 30 768 80 0 49 30 773 46 99 Zertifiziert nach ISO 9001 ISO 13485...

Отзывы: