background image

Sterilization for the US market

Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization.

Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave.

To achieve a sterility assurance level of 10

-6

, Aesculap recommends the following parameters:

*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an
Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles
may also be suitable, however individuals or hospitals not using the recommended method are advised to validate
any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility
after processing, such as a wrap, pouch, etc.

Storage

Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.

Technical Service

For service and repairs, please contact your national B. Braun/Aesculap agency.

Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and for-
feiture of applicable licenses.

Service addresses

Aesculap Technischer Service
Am Aesculap-Platz 
78532 Tuttlingen / Germany
Phone:

+49 (7461) 95-1602

Fax:

+49 (7461) 16-5621

E-Mail:

[email protected]

Or in the US: 
Attn. Aesculap Technical Services 
615 Lambert Pointe Drive 
Hazelwood 
MO, 63042 
Aesculap Repair Hotline 
Phone:

+1 (800) 214-3392 

Fax:

+1 (314) 895-4420

Other service addresses can be obtained from the address indicated above.

Disposal

Reprocess the product prior to disposal, see Validated reprocessing procedure.

Adhere to national regulations when disposing of or recycling the product, its components and its packaging!

Distributor in the US/Contact in Canada for product information and 
complaints

3773 Corporate Parkway 
Center Valley, PA, 18034,
USA

TA-Nr. 012787     02/13     V6     Änd.-Nr. 46013

Aesculap Orga Tray/Sterile container (perforated bottom)
Minimum cycle parameters*

Sterilization method

Temp.

Time

Minimum drying time

Prevacuum

270 °F/275 °F

4 min

20 min

Содержание Aesculap 012787

Страница 1: ... beschrijving Instructies voor de reiniging en sterilisatie van herbruikbare instrument Bruksanvisning Teknisk beskrivning Anvisningar om beredning av verktyg som kan återanvändas Инструкция по примению Техническое описание Указания по обработке инструмента многоразового использования Návod k použití Technický popis Pokyny k úpravě opětovně použitelných nástrojů Instrukcja użytkowania Opis technic...

Страница 2: ...ca tion process solution only needs to be of pH 8 to cause visible surface changes Material damage such as corrosion cracks fracturing premature aging or swelling Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion Further detailed advice on hygienically safe and material value preserving reprocessing can be found at www a k i org l...

Страница 3: ... solution at least five times using a disposable syringe Phase II Rinse flush the product thoroughly all accessible surfaces under running water Mobilize non rigid components such as set screws joints etc during rinsing Mechanical alkaline cleaning and thermal disinfecting Machine type single chamber cleaning disinfection device without ultrasound D W Drinking water FD W Fully desalinated water de...

Страница 4: ...al Service For service and repairs please contact your national B Braun Aesculap agency Modifications carried out on medical technical equipment may result in loss of guarantee warranty rights and for feiture of applicable licenses Service addresses Aesculap Technischer Service Am Aesculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1602 Fax 49 7461 16 5621 E Mail ats aesculap de Or in the US ...

Страница 5: ...ren Optische Materialveränderungen wie z B Verblassen oder Farbveränderungen bei Titan oder Aluminium Bei Alu minium können sichtbare Oberflächenveränderungen bereits bei einem pH Wert von 8 in der Anwendungs Gebrauchslösung auftreten Materialschäden wie z B Korrosion Risse Brüche vorzeitige Alterung oder Quellung Zur Reinigung keine Metallbürsten oder keine anderen die Oberfläche verletzenden Sch...

Страница 6: ...spritze gründlich durchspülen jedoch mindestens 5 mal Phase II Produkt vollständig alle zugänglichen Oberflächen unter fließendem Wasser ab durchspülen Nicht starre Komponenten wie z B Stellschrauben Gelenke etc bei der Spülung bewegen Maschinelle alkalische Reinigung und thermische Desinfektion Gerätetyp Einkammer Reinigungs Desinfektionsgerät ohne Ultraschall T W Trinkwasser VE W Vollentsalztes ...

Страница 7: ...ice Adressen Aesculap Technischer Service Am Aesculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1602 Fax 49 7461 16 5621 E Mail ats aesculap de Weitere Service Adressen erfahren Sie über die oben genannte Adresse Entsorgung Produkt vor der Entsorgung aufbereiten siehe Validiertes Aufbereitungsverfahren Bei Entsorgung oder Recycling des Produkts dessen Komponenten und deren Verpackung die nat...

Страница 8: ... cas contraire les problèmes suivants peuvent survenir Modification d aspect du matériau p ex palissement ou altérations de couleur du titane ou de l aluminium Sur l aluminium des altérations de surface visibles peuvent se produire dès une valeur de pH de 8 dans la solution utilisée Détériorations de matériau telles que corrosion fissures cassures vieillissement prématuré ou dilatations Pour le ne...

Страница 9: ... intégralement ces emplacements avec la solution de décontamination nettoyante à l aide d une seringue à usage unique adaptée et au moins à 5 reprises Phase II Laver rincer le produit intégralement toutes les surfaces accessibles sous l eau courante Pendant le rinçage faire bouger les composants non rigides tels que vis de réglage articulations etc Nettoyage alcalin en machine et décontamination t...

Страница 10: ...e perte des droits à garantie de même que d éventuelles autorisations Adresses de service Aesculap Technischer Service Am Aesculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1602 Fax 49 7461 16 5621 E Mail ats aesculap de Pour obtenir d autres adresses de service contactez l adresse ci dessus Elimination Traiter le produit avant l élimination voir Procédé de traitement stérile validé Lors de ...

Страница 11: ...VAH DGHM o la FDA o con marcado CE y recomendados por el fabricante en cuanto a su compatibilidad con el material Deberán cumplirse estrictamente todas las instrucciones del fabricante para el producto químico De lo contrario podrían surgir los siguientes problemas Alteraciones ópticas del material como decoloración o cambio de color en el caso del titanio o del aluminio En las superficies de alum...

Страница 12: ...ucto con un cepillo de limpieza adecuado hasta eliminar todos los restos de la superficie Cuando proceda deberá cepillarse las superficies no visibles con un cepillo adecuado durante al menos 1 min Durante la limpieza mover los componentes móviles como p ej tornillos de ajuste articulaciones etc A continuación lavar profusamente estos puntos con la solución desinfectante con acción limpiadora como...

Страница 13: ...egido contra el polvo y a temperatura constante Servicio de Asistencia Técnica Para asistencia técnica y reparaciones diríjase a su distribuidor nacional de B Braun Aesculap Si se realizan modificaciones en el equipo médico técnico se extinguirá la garantía y el derecho de garantía así como las posibles homologaciones Direcciones de la Asistencia Técnica Aesculap Technischer Service Am Aesculap Pl...

Страница 14: ...re di sostanze chimiche Altrimenti possono emergere i seguenti problemi Alterazioni ottiche dei materiali come ad es scoloriture o alterazioni cromatiche per il titanio o l alluminio Per l alluminio alterazioni superficiali visibili possono verificarsi già a partire da un valore pH 8 della soluzione d uso Danni materiali come ad es corrosione crepe rotture invecchiamento precoce o rigonfiamenti Pe...

Страница 15: ...empre per almeno per 5 volte Fase II Sciacquare completamente il prodotto tutte le superfici accessibili sotto l acqua corrente Durante il risciacquo muovere i componenti non rigidi come ad es viti di arresto snodi ecc Pulizia automatica alcalina e disinfezione termica Modello di apparecchio Lavatrice disinfettore monocamera senza ultrasuoni A P Acqua potabile A CD Acqua completamente desalinizzat...

Страница 16: ... delle omologazioni Indirizzi dei centri assistenza Aesculap Technischer Service Am Aesculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1602 Fax 49 7461 16 5621 E Mail ats aesculap de Gli altri indirizzi dell assistenza possono essere richiesti all indirizzo predetto Smaltimento Prima dello smaltimento eseguire la preparazione sterile del prodotto vedere Procedimento di preparazione ste rile ...

Страница 17: ...químicos Caso con trário poderão surgir os seguintes problemas Alterações ópticas do material por exemplo desbotamento ou alterações de cor no titânio ou alumínio No caso do alumínio podem ocorrer alterações visíveis da superfície mesmo em soluções de aplicação utilização com um valor de pH 8 Danos no material por exemplo corrosão fendas rupturas desgaste prematuro ou dilatação Para a limpeza não ...

Страница 18: ...e Durante a lavagem mover os componentes móveis como por exemplo parafusos de ajuste articulações etc Limpeza alcalina automática e desinfecção térmica Tipo de aparelho aparelho de limpeza desinfecção de câmara única sem ultra sons A P Água potável A CD Água completamente dessalinizada desmineralizada em termos microbiológicos no mínimo com qualidade de água potável Recomenda se BBraun Helimatic C...

Страница 19: ...tia e responsabilidade do fabricante bem como de possíveis licenças Endereços de assistência técnica Aesculap Technischer Service Am Aesculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1602 Fax 49 7461 16 5621 E Mail ats aesculap de Pode obter outros endereços de assistência técnica através do endereço acima referido Eliminação Preparar o produto antes de proceder à sua eliminação ver Método ...

Страница 20: ...richtlijnen van de chemicaliënfabrikant moeten strikt worden nage leefd In het andere geval kan dit tot de volgende problemen leiden Optische verandering van het materiaal bijv verbleken of kleurverandering van titanium of aluminium Bij alu minium kunnen zichtbare oppervlakteveranderingen reeds optreden bij een pH waarde vanaf 8 in de gebruiks oplossing Materiële schade zoals corrosie scheurtjes b...

Страница 21: ... product volledig alle bereikbare oppervlakken onder stromend water af door Beweeg tijdens het spoelen alle niet starre delen zoals stelschroeven scharnieren enz Machinale alkalische reiniging en thermische desinfectie Type apparaat enkele kamer reinigings desinfectieapparaat zonder ultrasoonreiniging D W Drinkwater DM W Gedemineraliseerd water microbiologisch minstens drinkwaterkwaliteit Aanbevol...

Страница 22: ...Service Am Aesculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1602 Fax 49 7461 16 5621 E Mail ats aesculap de Andere service adressen zijn verkrijgbaar op het bovengenoemde adres Verwijdering Reinig en steriliseer het product voorafgaande aan de verwijdering ervan zie Gevalideerd reinigings en desin fectieproces De verwijdering of recycling van het product de onderdelen en verpakking hiervan...

Страница 23: ...kemikalietillverkaren när det gäller materialkompabilitet Samtliga användningsföreskrifter från kemikalietillverkaren ska efterföljas strikt I annat fall kan följande problem uppstå Optiska förändringar av materialet som t ex blekning eller färgförändringar på titan eller aluminium På alumi nium kan synliga ytförändringar uppträda redan vid pH värde 8 i användnings brukslösningen Skador på materia...

Страница 24: ...er etc vid sköljningen Maskinell alkalisk rengöring och termisk desinficering Apparattyp Rengörings desinfektionsapparat med en kammare utan ultraljud DV Dricksvatten TAV Totalt avsaltat vatten avjoniserat mikrobiologiskt av minst dricksvattenkvalitet Rekommendation BBraun Helimatic Cleaner alcaline Kontrollera efter den maskinella rengöringen desinficeringen att det inte finns rester på ytor som ...

Страница 25: ...p Technischer Service Am Aesculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1602 Fax 49 7461 16 5621 E Mail ats aesculap de Ytterligare service adresser kan fås via ovan nämnda adress Avfallshantering Bered produkten före avfallshanteringen se Validerad beredningsmetod De nationella föreskrifterna måste ovillkorligen följas vid kassering eller återvinning av produkten dess kompo nenter eller...

Страница 26: ...й лазером надписи на нержавеющей стали что сделает невозмож ным ее прочтение визуально или машинным способом Под воздействием хлора или хлорсодержащих остатков содержащихся например в загрязнениях оставшихся после операции в лекарствах растворах поваренной соли в воде используемой для очистки дезинфекции и стерилизации на нержавеющей стали могут возникнуть очаги коррозии точечная коррозия коррозия...

Страница 27: ...е зафиксированы неподвижно Дать стечь остаткам воды Фаза III Полностью погрузить продукт в дезинфицирующий раствор При дезинфекции сдвигать детали например регулировочные винты шарниры и т д которые не зафиксированы неподвижно В начале обработки промыть внутренние просветы соответствующим одноразовым шприцем не менее 5 раз При этом следить за тем чтобы все доступные поверхности были смочены Фаза I...

Страница 28: ...ции например в контейнер Aesculap для стерилизации Убедиться в том что упаковка защищает от повторного загрязнения изделие во время хранения Стерилизация паром Убедиться в том что стерилизующий состав имеет доступ ко всем внешним и внутренним поверхно стям например открыв вентили и краны Утвержденный метод стерилизации Стерилизация паром методом дробной вакуумной стерилизации Использовать паровой ...

Страница 29: ...výrobce chemikálie je nutno důsledně dodržovat V opačném případě mohou nastat následující problémy Optické změny materiálu jako např vyblednutí nebo změny barvy u titanu a hliníku U hliníku může dojít k viditelným změnám na povrchu již při hodnotě pH 8 aplikačního hotového roztoku Poškození materiálu jako např koroze trhlinky nalomení předčasné stárnutí nebo bobtnání K čištění nepoužívejte kovové ...

Страница 30: ...ěcími šrouby klouby atd v průběhu proplachování pohybujte Strojní alkalické čištění a tepelná desinfekce Typ přístroje Jednokomorový čistící desinfekční přístroj bez ultrazvuku PV Pitná voda DEV Zcela solí zbavená voda demineralizovaná z mikrobiologického hlediska minimálně v kvalitě pitné vody Doporučen BBraun Helimatic Cleaner alcaline Po strojovém čištění a desinfekci zkontrolujte všechny vidit...

Страница 31: ...sů se dozvíte prostřednictvím výše uvedené adresy Likvidace Výrobek před likvidací upravte viz Validovaná metoda úpravy Při likvidaci nebo recyklaci výrobku jeho komponent a jejich obalů dodržujte národní předpisy Distributor B BRAUN Medical s r o V Parku 2335 20 148 00 Praha 4 Tel 271 091 111 Fax 271 091 112 E mail servis cz bbraun com TA Nr 012787 02 13 V6 Änd Nr 46013 ...

Страница 32: ... przez producenta środków chemicznych W prze ciwnym razie może to spowodować następujące problemy Wizualne zmiany produktu jak np wyblaknięcie lub przebarwienia tytanu lub aluminium W przypadku alumi nium widoczne zmiany na powierzchni mogą wystąpić już wówczas gdy pH roztworu roboczego stosowanego wynosi 8 Szkody takie jak korozja rysy złamania przedwczesne starzenie się materiału lub pęcznienie ...

Страница 33: ...otyczy niewidzialne powierzchnie należy co najmniej 1 przeczyścić odpowiednią szczotką do czyszczenia Podczas czyszczenia należy poruszać ruchomymi elementami urządzeń takimi jak śruby regulacyjne przeguby itp Następnie miejsca te należy dokładnie przepłukać co najmniej 5 razy aktywnie czyszczącym roztworem dezyn fekcyjnym za pomocą odpowiedniej strzykawki jednorazowej Faza II Produkt należy całko...

Страница 34: ...ównanej temperaturze Serwis techniczny W sprawie konkretnych usług serwisowych proszę się skontaktować z właściwym dla Państwa krajowym przed stawicielstwem firmy B Braun Aesculap Wprowadzanie zmian konstrukcyjnych do urządzeń medycznych może skutkować utratą praw gwarancyjnych praw z tytułu rękojmi jak również istniejących dopuszczeń Adresy punktów serwisowych Aesculap Technischer Service Am Aesc...

Страница 35: ...kých látok sa musia prísne dodržiavať V ostatných prípadoch to môže viesť k nasledujúcim problémom Optické zmeny materiálu ako napr vyblednutie alebo zmena farby titánu a hliníka V prípade hliníka môžu nastať viditeľné zmeny povrchu už pri pH hodnote 8 v aplikovanom užívateľskom roztoku Materiálne škody ako je napr korózia trhliny lomy predčasné stárnutie alebo napúčanie Na čistenie nepoužívajte ž...

Страница 36: ...najmenej 5 krát Fáza II Celý nástroj všetky prístupné povrchy opláchnite prepláchnite pod tečúcou vodou Pohyblivými komponentmi ako sú napr nastaviteľné skrutky kĺby atď počas oplachovania pohybujte Strojové alkalické čistenie a tepelná dezinfekcia Typ prístroja Jednokomorový čistiaci dezinfekčný prístroj bez ultrazvuku PV Pitná voda DV Voda úplne zbavená solí demineralizovaná voda mikrobiologicky...

Страница 37: ...esculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1602 Fax 49 7461 16 5621 E Mail ats aesculap de Ďalšie servisné adresy získate na hore uvedenej adrese Likvidácia Výrobok pred likvidovaním upravte pozri Validované postupy prípravy Pri likvidácii alebo recyklácii výrobku obsahujú jeho zložky a obal národné predpisy Distribútor B BRAUN Medical s r o Handlovská 19 Bratislava 851 01 Slovensko T...

Страница 38: ...madde üreticisinin tüm uygulamaları sıkı sıkıya müşahade edilmelidir Aksi halde bunun sonucunda aşağıda sayılan problemler ortaya çıka bilir Malzemede görüntü değişimleri örn solma ya da titan ya da alüminyumda renk değişikliği Alüminyumda pH değeri 8 den itibaren uygulama kullanım solüsyonunda görünür yüzey değişimleri ortaya çıkabilir Malzeme hasarları örn korozyon çatlaklar kırıklar erken yaşla...

Страница 39: ... olmamak üzere iyice durulayınız Evre II Ürünü tamamıyla bütün erişilebilir yüzeyleri akan musluk suyu altında iyice yıkayınız durulayınız Sabit olmayan komponentleri örn ayar vidaları eklemler vs yıkama sırasında hareket ettiriniz Makineyle alkalik temizlik ve termik dezenfeksiyon Cihaz tipi Ultrasonsuz tek bölmeli temizlik dezenfeksiyon cihazı İS İçme suyu TTAS Tuzdan tamamen arındırılmış su dem...

Страница 40: ...esculap Technischer Service Am Aesculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1602 Fax 49 7461 16 5621 E Mail ats aesculap de Diğer servis adreslerini yukarıda yazılı adresten öğrenebilirsiniz Atık bertarafı Ürünü tasfiye etmeden önce hazırlayın bkz Validasyonu yapılmış hazırlama yöntemi Ürünün komponentlerinin ve ambalajının atık bertarafı ya da geri dönüşümü için mutlaka ülkenizdeki ku...

Отзывы: